Scoville J R
Regulatory Affairs and Quality Assurance, Englewood, Colorado, USA.
Compendium. 1994 Dec;15(12):1472-4, 1476, 1478; quiz 1478.
The use of at least two biologic indicators to monitor a sterilization process greatly enhances failure detection. A retrospective study of a sterilization monitoring service's database revealed that 48.7% of tests using two test strips were single-strip positive. This indicates that there is a nonuniformity in the sterilization conditions within the sterilizer chamber and that there is a need for at least two test strips to detect this nonuniformity.
使用至少两个生物指示剂来监测灭菌过程可大大提高故障检测率。一项对灭菌监测服务数据库的回顾性研究表明,使用两条测试条的测试中有48.7%为单条阳性。这表明灭菌器腔内的灭菌条件存在不均匀性,并且需要至少两条测试条来检测这种不均匀性。