Van Loan M D, Kopp L E, King J C, Wong W W, Mayclin P L
United States Department of Agriculture, Western Human Nutrition Research Center, Presidio of San Francisco, California, 94129, USA.
J Appl Physiol (1985). 1995 Mar;78(3):1037-42. doi: 10.1152/jappl.1995.78.3.1037.
The increase in body water during pregnancy is responsible for the largest portion of weight gain and is of interest of clinical practitioners. However, assessing changes in body fluids is not easily accomplished during pregnancy. The purpose of this study was to examine the accuracy of bioelectrical impedance spectroscopy for estimating fluid volumes before, during, and after pregnancy. Ten healthy adult women were recruited for the study. Total body water (TBW) and extracellular fluid (ECF) volume were measured at baseline (preconception); 8-10, 24-26, and 34-36 wk of gestation; and 4-6 wk postpartum by deuterium oxide and NaBr dilution, respectively. Estimates of TBW and ECF were also obtained by bioimpedance spectroscopy (BIS). At baseline, mean values for dilution and BIS estimates of TBW and ECF were 33.2 +/- 4.6 (SD) vs. 31.6 +/- 6.2 liters and 15.2 +/- 1.3 vs. 16.9 +/- 2.3 liters, respectively. TBW and ECF estimated by BIS were not significantly different from the dilution values at any time point. These results suggest that BIS may be useful in estimating volumes of ECF and TBW during pregnancy.
孕期体重增加的最大部分是由身体水分增加引起的,这是临床医生所关注的。然而,孕期评估体液变化并非易事。本研究的目的是检验生物电阻抗光谱法在孕期前、孕期中和产后估计液体量的准确性。招募了10名健康成年女性参与该研究。分别在基线期(受孕前)、妊娠8 - 10周、24 - 26周、34 - 36周以及产后4 - 6周,通过氧化氘和溴化钠稀释法测量总体水(TBW)和细胞外液(ECF)量。还通过生物电阻抗光谱法(BIS)获得TBW和ECF的估计值。在基线期,TBW和ECF的稀释法测量值与BIS估计值的平均值分别为33.2±4.6(标准差)升对31.6±6.2升,以及15.2±1.3升对16.9±2.3升。在任何时间点,BIS估计的TBW和ECF与稀释法测量值均无显著差异。这些结果表明,BIS可能有助于估计孕期的ECF和TBW量。