Validation of right ventricular volume measurements by magnetic resonance imaging in small hearts using a fetal lamb model.

作者信息

Stauffer N R, Greenberg S B, Marks L A, Singh G K, Siderio D L

机构信息

Pennsylvania Hospital, Philadelphia, USA.

出版信息

Invest Radiol. 1995 Feb;30(2):87-9. doi: 10.1097/00004424-199502000-00004.

Abstract

RATIONALE AND OBJECTIVES

Cine-magnetic resonance (MR) imaging is established as an accurate method of determining cardiac ventricular volumes in adults but has not been validated for the smaller ventricular volumes found in infants. The purpose of the authors study was to evaluate the accuracy and reproducibility of MR imaging to measure small ventricle volumes.

METHODS

The volume of silicone-rubber casts of the right ventricle of 11 fetal lamb hearts was measured by MR imaging using different imaging planes and slice section thicknesses and by water displacement.

RESULTS

The ventricle volumes of the 11 hearts ranged from 0.45 to 3.00 mL. Accurate and reproducible values for the right ventricle volumes were obtained by MR imaging. Thin sections (3 mm) in the axial plane produced the best results.

CONCLUSIONS

Based on studies in the nonmoving heart, MR imaging can accurately measure the small ventricular volumes expected in infants.

摘要

原理与目的

电影磁共振(MR)成像已被确立为一种测定成人心室容积的准确方法,但对于婴儿较小的心室容积尚未得到验证。作者研究的目的是评估MR成像测量小心室容积的准确性和可重复性。

方法

使用不同的成像平面和切片厚度,通过MR成像和水置换法测量11个胎羊心脏右心室的硅橡胶铸型的容积。

结果

11个心脏的心室容积范围为0.45至3.00毫升。通过MR成像获得了右心室容积的准确且可重复的值。轴向平面的薄切片(3毫米)产生了最佳结果。

结论

基于对静止心脏的研究,MR成像可以准确测量婴儿预期的小心室容积。