Hess C F, Kortmann R D, Jany R, Hamberger A, Bamberg M
Department of Radiotherapy, University of Tübingen, Germany.
Radiother Oncol. 1995 Jan;34(1):69-72. doi: 10.1016/0167-8140(94)01497-q.
Deviations between simulation and first check films were quantitatively assessed for 95 unselected head and neck cancer patients. All measured deviations--calculated on the basis of a total of 190 simulation and 380 verification films--were normally distributed, with mean values of 0-3 mm and standard deviations of 3-5 mm. Of the absolute deviations, 50% and 95% were within 3 mm and 9 mm, respectively. These results should be considered in clinical practice when prescribing safety margins and adequate cut off doses for sparing critical organs in head and neck cancer.
对95例未经挑选的头颈癌患者的模拟影像与首次检查影像之间的偏差进行了定量评估。所有测量的偏差——基于总共190幅模拟影像和380幅验证影像计算得出——呈正态分布,平均值为0至3毫米,标准差为3至5毫米。在绝对偏差中,50%和95%分别在3毫米和9毫米以内。在对头颈癌关键器官进行安全边缘设定和确定合适的剂量截止值时,临床实践中应考虑这些结果。