[Dynamic 3D MR mammography using a rapid gradient-echo sequence].

作者信息

Müller-Schimpfle M, Rieber A, Kurz S, Stern W, Claussen C D

机构信息

Radiologische Klinik, Abteilung Radiologische Diagnostik, Universität Tübingen.

出版信息

Rofo. 1995 Jan;162(1):13-9. doi: 10.1055/s-2007-1015827.

Abstract

Up to now, MR mammography is performed by two different methods: 2-D techniques with high temporal resolution and 3-D techniques with high spatial resolution. This article investigates whether a dynamic examination of the breast using a novel 3-D-gradient-echo sequence is feasible with sufficient spatial and temporal resolution. MR imaging was performed on a 1.0 Tesla imager using a gradient field strength of 15 mT/m. Phantom measurements were done to evaluate the correlation between signal intensity and contrast medium concentration for different sequences. Subsequently, 40 patients with 22 histologically verified lesions were examined using the double breast coil and a novel FLASH3D sequence (TR/TE/fa 9/3/50 degrees). Reading of the films in standardized documentation, multiplanar reconstruction (MPR), calculation and maximum intensity projection (MIP) of subtraction images as well as signal-to-time curve calculations in selected ROIs were performed for data evaluation. In the phantom measurements the FLASH3D-9/3/50 degrees sequence yielded the best correlation between signal intensity and Gd-DTPA concentration. The sequence provided good visualization even of small lesions. The 3-D postprocessing procedures facilitated the detection and localization of the lesions. Therefore, the new FLASH3D-9/3/50 degrees sequence enables a dynamic 3-D examination of the breast with a sufficient spatial and temporal resolution.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验