Pulse sequence design for MR velocity mapping of complex flow: notes on the necessity of low echo times.

作者信息

Ståhlberg F, Thomsen C, Söndergaard L, Henriksen O

机构信息

Danish Center of Magnetic Resonance, Hvidovre Hospital, University of Copenhagen.

出版信息

Magn Reson Imaging. 1994;12(8):1255-62. doi: 10.1016/0730-725x(94)90090-e.

Abstract

Lowering of the echo time (TE) has been proposed as a way to reduce effects of phase dispersion in MR velocity mapping, because a low TE reduces sensitivity to higher-order motion terms while first-order velocity sensitivity is maintained. Methods of lowering TE involves the use of extreme gradient ramp times and gradient strengths as well as reduction of the duration of transmit/receive windows, the latter method causing decrements in image resolution. When reducing higher-order sensitivity, however, it is not the overall TE that is the critical parameter, but rather the time pattern of the gradients used in the experiment. Hence, changes in TE without subsequent variations in gradient pattern would, according to theory, not affect quantitative measurements of complex flow and vice versa. In this study, we experimentally demonstrate this relation and utilize the experience to create a sequence robust towards complex flow without sacrifices in image resolution. Our experimental observations show that variations in TE alone while maintaining the time course of the velocity-encoding gradient does not significantly affect measurements of through-plane average complex flow in the studied velocity range. A parameter that cannot be measured as accurately if TE is increased is the peak flow. A phase mapping sequence with prolonged TE from 3 ms to 5 ms but with short duration of the velocity-encoding (section-selective) gradient and improved in-plane resolution was demonstrated in vivo.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验