使用眼压笔测量带状角膜病变或角膜粘连患者的眼压。

Tono-Pen determination of intraocular pressure in patients with band keratopathy or glued cornea.

作者信息

Azuara-Blanco A, Bhojani T K, Sarhan A R, Pillai C T, Dua H S

机构信息

Department of Ophthalmology, University Hospital, Queen's Medical Centre, University of Nottingham.

出版信息

Br J Ophthalmol. 1998 Jun;82(6):634-6. doi: 10.1136/bjo.82.6.634.

Abstract

AIMS

To evaluate the intraocular pressure (IOP) measurements in patients with band keratopathy or glued corneas obtained from affected and non-affected areas.

METHODS

15 patients with band keratopathy or glued corneas were prospectively recruited. When both eyes were affected, only the right eye was analysed. Tono-Pen readings of IOP were obtained sequentially from the affected and non-affected corneal surface. Additionally, Goldmann applanation tonometry was attempted.

RESULTS

Determination of IOP with the Tono-Pen was possible in all cases, while Goldmann tonometry was not performed in three patients because of severe corneal irregularities. The average of the Tono-Pen readings obtained from the affected cornea (34.8 (SD 14.0) mm Hg) was consistently and significantly higher (p < 0.001) than mean IOP obtained by the Tono-Pen from the non-affected area (14.8 (4.3) mm Hg). The average of Goldmann tonometry readings (14.4 (6.1) mm Hg) did not differ significantly from the Tono-Pen values obtained from the non-affected corneal area (p = 0.47) but was significantly lower than the Tono-Pen measurements obtained from the affected area (p < 0.001).

CONCLUSION

In patients with band keratopathy or glued corneas determination of IOP by Tono-Pen tonometry varies from affected to non-affected area. The Tono-Pen overestimates the level of IOP when it is applied to areas with band keratopathy or with glue.

摘要

目的

评估带状角膜病变或角膜粘连患者患眼及未患眼区域的眼压测量值。

方法

前瞻性招募了15例带状角膜病变或角膜粘连患者。若双眼均受累,则仅分析右眼。依次从患眼和未患眼的角膜表面获取眼压计(Tono-Pen)测量的眼压读数。此外,尝试进行Goldmann压平眼压测量。

结果

所有病例均可用眼压计测定眼压,3例患者因角膜严重不规则而未进行Goldmann眼压测量。从患眼角膜获得的眼压计读数平均值(34.8(标准差14.0)mmHg)始终显著高于(p<0.001)从未患眼区域获得的眼压计平均眼压(14.8(4.3)mmHg)。Goldmann眼压测量读数平均值(14.4(6.1)mmHg)与从未患眼角膜区域获得的眼压计值无显著差异(p=0.47),但显著低于从患眼区域获得的眼压计测量值(p<0.001)。

结论

对于带状角膜病变或角膜粘连患者,用眼压计测量眼压时,患眼与未患眼区域的测量值不同。将眼压计应用于带状角膜病变或有胶水的区域时,会高估眼压水平。

相似文献

引用本文的文献

3
Limbal rebound tonometry: clinical comparisons and applications.角膜缘回弹眼压测量法:临床比较与应用
Graefes Arch Clin Exp Ophthalmol. 2017 Sep;255(9):1795-1799. doi: 10.1007/s00417-017-3725-7. Epub 2017 Jun 28.
5
Detecting IOP Fluctuations in Glaucoma Patients.检测青光眼患者的眼压波动
Open Ophthalmol J. 2016 Feb 29;10:44-55. doi: 10.2174/1874364101610010044. eCollection 2016.

本文引用的文献

2
The Goldmann applanation tonometer.戈德曼压平眼压计。
Am J Ophthalmol. 1958 Dec;46(6):865-9. doi: 10.1016/0002-9394(58)90998-x.
5
Assessment of intraocular pressure by palpation.
Am J Ophthalmol. 1995 May;119(5):650-1. doi: 10.1016/s0002-9394(14)70227-2.
8
Clinical evaluation of the Oculab Tono-Pen.Oculab眼压笔的临床评估
Am J Ophthalmol. 1987 Aug 15;104(2):168-73. doi: 10.1016/0002-9394(87)90010-9.
10
Comparison of the Tono-Pen to the Goldmann applanation tonometer.
Arch Ophthalmol. 1988 Jun;106(6):750-3. doi: 10.1001/archopht.1988.01060130820030.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验