Azuara-Blanco A, Bhojani T K, Sarhan A R, Pillai C T, Dua H S
Department of Ophthalmology, University Hospital, Queen's Medical Centre, University of Nottingham.
Br J Ophthalmol. 1998 Jun;82(6):634-6. doi: 10.1136/bjo.82.6.634.
To evaluate the intraocular pressure (IOP) measurements in patients with band keratopathy or glued corneas obtained from affected and non-affected areas.
15 patients with band keratopathy or glued corneas were prospectively recruited. When both eyes were affected, only the right eye was analysed. Tono-Pen readings of IOP were obtained sequentially from the affected and non-affected corneal surface. Additionally, Goldmann applanation tonometry was attempted.
Determination of IOP with the Tono-Pen was possible in all cases, while Goldmann tonometry was not performed in three patients because of severe corneal irregularities. The average of the Tono-Pen readings obtained from the affected cornea (34.8 (SD 14.0) mm Hg) was consistently and significantly higher (p < 0.001) than mean IOP obtained by the Tono-Pen from the non-affected area (14.8 (4.3) mm Hg). The average of Goldmann tonometry readings (14.4 (6.1) mm Hg) did not differ significantly from the Tono-Pen values obtained from the non-affected corneal area (p = 0.47) but was significantly lower than the Tono-Pen measurements obtained from the affected area (p < 0.001).
In patients with band keratopathy or glued corneas determination of IOP by Tono-Pen tonometry varies from affected to non-affected area. The Tono-Pen overestimates the level of IOP when it is applied to areas with band keratopathy or with glue.
评估带状角膜病变或角膜粘连患者患眼及未患眼区域的眼压测量值。
前瞻性招募了15例带状角膜病变或角膜粘连患者。若双眼均受累,则仅分析右眼。依次从患眼和未患眼的角膜表面获取眼压计(Tono-Pen)测量的眼压读数。此外,尝试进行Goldmann压平眼压测量。
所有病例均可用眼压计测定眼压,3例患者因角膜严重不规则而未进行Goldmann眼压测量。从患眼角膜获得的眼压计读数平均值(34.8(标准差14.0)mmHg)始终显著高于(p<0.001)从未患眼区域获得的眼压计平均眼压(14.8(4.3)mmHg)。Goldmann眼压测量读数平均值(14.4(6.1)mmHg)与从未患眼角膜区域获得的眼压计值无显著差异(p=0.47),但显著低于从患眼区域获得的眼压计测量值(p<0.001)。
对于带状角膜病变或角膜粘连患者,用眼压计测量眼压时,患眼与未患眼区域的测量值不同。将眼压计应用于带状角膜病变或有胶水的区域时,会高估眼压水平。