Suppr超能文献

无晶状体密度检测的短波自动视野计

Short-wavelength automated perimetry without lens density testing.

作者信息

Sample P A, Martinez G A, Weinreb R N

机构信息

Glaucoma Center and Visual Function Laboratory, University of California at San Diego, La Jolla 92093-0946.

出版信息

Am J Ophthalmol. 1994 Nov 15;118(5):632-41. doi: 10.1016/s0002-9394(14)76578-x.

Abstract

Three experiments determined whether an estimate of localized short-wavelength visual field loss could provide enough specificity and sensitivity for glaucoma diagnosis, without correcting the fields for lens density. We used 100 normal, 53 glaucoma suspect, and 95 glaucoma eyes (or a subset of these, where noted). The first two experiments showed that lens density has a diffuse effect on the visual field, which is more linear than the effect of age. The third experiment showed that, with the glaucoma hemifield test, comparable results were obtained for short-wavelength fields either corrected or uncorrected for lens density (sensitivities of 81% and 76% and specificities of 82% and 90%, respectively). We conclude that it is possible to eliminate the lens density measurement with this approach, thereby reducing test time by 40 minutes. This should improve substantially the clinical utility of short-wavelength automated perimetry.

摘要

三项实验确定了在不针对晶状体密度校正视野的情况下,局部短波长视野缺损的估计值是否能为青光眼诊断提供足够的特异性和敏感性。我们使用了100只正常眼睛、53只疑似青光眼眼睛和95只青光眼眼睛(或视情况使用这些眼睛的子集)。前两项实验表明,晶状体密度对视野有弥散性影响,这种影响比年龄的影响更呈线性。第三项实验表明,使用青光眼半视野检测法,无论是否针对晶状体密度校正短波长视野,均可获得可比结果(敏感性分别为81%和76%,特异性分别为82%和90%)。我们得出结论,采用这种方法可以省去晶状体密度测量,从而将检测时间缩短40分钟。这将大幅提高短波长自动视野计的临床实用性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验