Brezina A, Schwaighofer B W
Abteilung für bildgebende Diagnostik, Evangelisches Krankenhaus Wien-Währing.
Wien Klin Wochenschr. 1994;106(18):584-9.
Over the past few years MRI has become important in the evaluation of inconclusive mammograms. These studies were performed on high-field strength machines, which have the disadvantage of limited accessibility. We evaluated 185 investigations on 169 patients on a 0.5 Tesla MR machine using a dynamic sequence and gadolinium (Gd) DTPA. The results in 107 patients were correlated with the histological findings and MRI assessment proved correct in 44 of 47 benign lesions (93.6%) and 57 of 60 malignant lesions (95%). The 3 false positive results occurred in patients with mastitis, fibroadenoma and an ectatic vessel, respectively. Of the 3 false negative results 2 occurred in patients with microcarcinomas and were due to partial-volume artefacts, whilst the third was due to a technical error. These results compare with those reported in the literature using mainly high-field strength machines. Furthermore, unsuspected second tumours were diagnosed in the ipsilateral breast of 6 women and infiltration of the adjacent thoracic wall in 4 women, findings which had not been evident on mammography. Hence, our results indicate that a mild-field strength MR machine is a valuable tool not only for the non-invasive differentiation between benign and malignant lesions of the breast, but also for planning therapeutic strategy.
在过去几年中,磁共振成像(MRI)在评估诊断不明确的乳房X光片方面变得至关重要。这些研究是在高场强机器上进行的,其缺点是可及性有限。我们使用动态序列和钆(Gd)二乙三胺五乙酸(DTPA),在一台0.5特斯拉的磁共振机器上对169例患者进行了185次检查。107例患者的结果与组织学结果相关,MRI评估在47例良性病变中的44例(93.6%)以及60例恶性病变中的57例(95%)中被证明是正确的。3例假阳性结果分别出现在患有乳腺炎、纤维腺瘤和扩张血管的患者中。在3例假阴性结果中,2例出现在患有微癌的患者中,是由于部分容积伪影,而第三例是由于技术错误。这些结果与主要使用高场强机器的文献报道结果相当。此外,在6名女性的同侧乳房中诊断出了未被怀疑的第二肿瘤,在4名女性中发现了邻近胸壁的浸润,这些发现乳房X光片上并不明显。因此,我们的结果表明,低场强磁共振机器不仅是用于乳腺良恶性病变无创鉴别诊断的有价值工具,也是用于制定治疗策略的有价值工具。