[Comparative study of 24-hour ambulatory blood pressure monitors with standard zero and random zero sphygmomanometers].

作者信息

Tislér A, Barna I, Châtel R

机构信息

I. sz. Belgyógyászati Klinika.

出版信息

Orv Hetil. 1994 Jun 26;135(26):1415-9.

PMID:8028897
Abstract

The aim of the study was to evaluate the accuracy of the most widespread 24-hour ambulatory blood pressure monitor in Hungary. The test was based on simultaneous measurement on the same arm with the test device and standard zero or random zero sphygmomanometer in 100 patients. The difference between the blood pressure values measured by the test device and by the standard device was calculated in each case, and a relationship between this difference and the actual blood pressure of the patient was analysed. Actual blood pressure was considered as the average of the blood pressure measured by the test and that by the standard device. Regarding the diastolic values, the mean difference between the values obtained by the test device and the standard zero sphygmomanometer was -3.8 +/- 7.55 mm Hg (p > 0.05), and that between the test device and the random zero sphygmomanometer was -0.1 +/- 6.05 mm Hg (p > 0.05). This differences did not reach statistical significance. Regarding the systolic values, the difference showed significant positive correlation with the actual blood pressure level of the patient, meaning that at higher blood pressure values the difference between the data gained by the test and standard device is greater than at lower actual blood pressure levels. Finally, according to the criteria of the British Hypertension Society, our device could be graded "C" both for diastolic and systolic values against the standard zero sphygmomanometer, and "C" for systolic and "B" for diastolic blood pressure against the random zero sphygmomanometer.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验