An in vitro study of mechanical heart valve sound loudness.

作者信息

Erickson R L, Thulin L I, Richard G J

机构信息

CarboMedics, Inc., Austin, Texas 78752.

出版信息

J Heart Valve Dis. 1994 May;3(3):330-4.

PMID:8087274
Abstract

The purpose of the study was to measure the valve sound loudness of eight 25 mm aortic mechanical heart valve prostheses under controlled laboratory conditions in an isolated acoustic environment. A specially designed single-chamber pulse duplicator was constructed within an anechoic box that isolated the sounds of the valve from external noise. The valves were closed with a stimulated aortic pressure waveform of three different amplitudes. Valve sound loudness was measured with ISO-532/B (Zwicker method), which is based on a model of loudness perception that is better than dB(A), the traditional A-weighted sound pressure level. The average loudness for each valve was St. Jude Medical, 0.97 sones; CarboMedics, 0.98 sones; Sorin Biomedica Bicarbon, 1.08 sones; Edwards Tekna, 1.92 sones; Edwards Duromedics, 2.36 sones; Monostrut, 2.48 sones; Sorin Biomedica Carbocast, 3.17 sones; and Medtronic Hall, 3.57 sones. There was a positive trend between valve sound loudness and aortic drive pressure for most valves in the study.

摘要

该研究的目的是在隔离声学环境的受控实验室条件下,测量八个25毫米主动脉机械心脏瓣膜假体的瓣膜声音响度。在一个消声箱内构建了一个专门设计的单腔脉冲复制器,该复制器将瓣膜的声音与外部噪声隔离开来。使用三种不同幅度的模拟主动脉压力波形关闭瓣膜。瓣膜声音响度采用ISO-532/B(齐克纳方法)进行测量,该方法基于一种响度感知模型,比传统的A加权声压级dB(A)更好。每个瓣膜的平均响度分别为:圣犹达医疗公司,0.97宋;卡博梅迪克斯公司,0.98宋;索林生物医学公司碳酸氢盐瓣膜,1.08宋;爱德华兹泰克纳瓣膜,1.92宋;爱德华兹杜罗梅迪克斯瓣膜,2.36宋;单支柱瓣膜,2.48宋;索林生物医学公司碳铸瓣膜,3.17宋;以及美敦力霍尔瓣膜,3.57宋。在该研究中,大多数瓣膜的瓣膜声音响度与主动脉驱动压力之间存在正相关趋势。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验