Erickson R L, Thulin L I, Richard G J
CarboMedics, Inc., Austin, Texas 78752.
J Heart Valve Dis. 1994 May;3(3):330-4.
The purpose of the study was to measure the valve sound loudness of eight 25 mm aortic mechanical heart valve prostheses under controlled laboratory conditions in an isolated acoustic environment. A specially designed single-chamber pulse duplicator was constructed within an anechoic box that isolated the sounds of the valve from external noise. The valves were closed with a stimulated aortic pressure waveform of three different amplitudes. Valve sound loudness was measured with ISO-532/B (Zwicker method), which is based on a model of loudness perception that is better than dB(A), the traditional A-weighted sound pressure level. The average loudness for each valve was St. Jude Medical, 0.97 sones; CarboMedics, 0.98 sones; Sorin Biomedica Bicarbon, 1.08 sones; Edwards Tekna, 1.92 sones; Edwards Duromedics, 2.36 sones; Monostrut, 2.48 sones; Sorin Biomedica Carbocast, 3.17 sones; and Medtronic Hall, 3.57 sones. There was a positive trend between valve sound loudness and aortic drive pressure for most valves in the study.
该研究的目的是在隔离声学环境的受控实验室条件下,测量八个25毫米主动脉机械心脏瓣膜假体的瓣膜声音响度。在一个消声箱内构建了一个专门设计的单腔脉冲复制器,该复制器将瓣膜的声音与外部噪声隔离开来。使用三种不同幅度的模拟主动脉压力波形关闭瓣膜。瓣膜声音响度采用ISO-532/B(齐克纳方法)进行测量,该方法基于一种响度感知模型,比传统的A加权声压级dB(A)更好。每个瓣膜的平均响度分别为:圣犹达医疗公司,0.97宋;卡博梅迪克斯公司,0.98宋;索林生物医学公司碳酸氢盐瓣膜,1.08宋;爱德华兹泰克纳瓣膜,1.92宋;爱德华兹杜罗梅迪克斯瓣膜,2.36宋;单支柱瓣膜,2.48宋;索林生物医学公司碳铸瓣膜,3.17宋;以及美敦力霍尔瓣膜,3.57宋。在该研究中,大多数瓣膜的瓣膜声音响度与主动脉驱动压力之间存在正相关趋势。