使用微型传感器进行肛门直肠测压。

Anorectal manometry with a microtransducer.

作者信息

Sundblad M, Hallböök O, Sjödahl R

机构信息

Department of Medico-Surgical Gastroenterology, Colorectal Surgery, Linköping University Hospital, Sweden.

出版信息

Eur J Surg. 1993 Jun-Jul;159(6-7):365-70.

PMID:8104499
Abstract

OBJECTIVE

To find out whether the direction of the sensor of the microtransducer or the position of the subject influenced anorectal manometric pressure recordings; to record resting and squeeze pressures and length of anal canal in five groups divided according to age, sex, and parity, and to study rectal sensitivity and response of intrarectal pressure to rectal distension.

DESIGN

Open study.

SETTING

University Hospital, Linköping, Sweden.

SUBJECTS

87 healthy volunteers with no signs or symptoms of anorectal disease.

INTERVENTIONS

Anorectal manometry with a microtransducer.

RESULTS

Neither the direction of the sensor nor the position of the subject influenced the recordings in five men and four women whose ages ranged from 40-58 years. In 60 further subjects grouped according to age, sex, and parity there was no difference in mean (SD) resting pressure between the groups. Maximal squeeze pressure was significantly greater among men than women, and there was no significant difference between parous and nulliparous women. The voluntary contractile force of the external sphincter was greater in young than older people, and in men compared with women. The functional length of the anal canal was greatest in older men and least in young parous women. In a further 18 volunteers aged 20-86 years rectal distension was felt at 25 (12) ml, and maximal tolerated volume was 202 (75) ml. Mean threshold volume required to release the rectosphincteric inhibitory reflex was 17 (8) ml.

CONCLUSIONS

Neither the direction of the sensor nor the position of the patient influenced the recordings. Resting pressure was the same irrespective of sex, age, or childbirth, but maximal squeeze pressure was significantly higher among men and those less than 50 years old.

摘要

目的

探究微型传感器的方向或受试者的体位是否会影响肛肠测压压力记录;记录根据年龄、性别和生育情况划分的五组人群的静息压力、收缩压力及肛管长度,并研究直肠敏感性和直肠扩张时直肠内压力的反应。

设计

开放性研究。

地点

瑞典林雪平大学医院。

受试者

87名无肛肠疾病体征或症状的健康志愿者。

干预措施

使用微型传感器进行肛肠测压。

结果

在年龄为40 - 58岁的5名男性和4名女性中,传感器方向和受试者体位均未影响记录结果。在另外60名根据年龄、性别和生育情况分组的受试者中,各组间平均(标准差)静息压力无差异。男性的最大收缩压力显著高于女性,经产妇和未产妇之间无显著差异。外括约肌的自主收缩力在年轻人中比老年人更大,在男性中比女性更大。肛管的功能长度在老年男性中最长,在年轻经产妇中最短。在另外18名年龄为20 - 86岁的志愿者中,直肠扩张在25(12)ml时被感知到,最大耐受容量为202(75)ml。释放直肠括约肌抑制反射所需的平均阈值容量为17(8)ml。

结论

传感器方向和患者体位均未影响记录结果。静息压力不受性别、年龄或生育情况的影响,但最大收缩压力在男性和50岁以下人群中显著更高。