Suppr超能文献

单软管和双软管系统的无创血压测量——混合气体有影响吗?

Noninvasive blood pressure measurements with single and twin-hose systems--do mixtures matter?

作者信息

Amoore J N, Scott D H

机构信息

Department of Medical Physics and Medical Engineering, Royal Infirmary of Edinburgh.

出版信息

Anaesthesia. 1993 Sep;48(9):799-802. doi: 10.1111/j.1365-2044.1993.tb07596.x.

Abstract

Noninvasive blood pressure monitors which use twin- and single-hose cuffs complicate the movement of patients between bed spaces with different cuff-hose systems. We assessed, using a commercially available noninvasive blood pressure monitor function analyser, the use of single-hose cuffs coupled via a Y-piece adaptor to the twin-hose of noninvasive monitors designed for use with twin-hose cuffs. Assessments were made at three simulated pressures (200/150 mmHg, 120/80 mmHg, and 60/30 mmHg) and at normal and weak arterial pulsations. The use of a single-hose cuff attached with the Y-piece adaptor did not adversely affect the performance of the monitors.

摘要

使用双管和单管袖带的无创血压监测仪会使患者在配备不同袖带-软管系统的床位之间移动变得复杂。我们使用市售的无创血压监测仪功能分析仪,评估了通过Y形接头适配器连接到专为双管袖带设计的无创监测仪双管上的单管袖带的使用情况。在三种模拟压力(200/150 mmHg、120/80 mmHg和60/30 mmHg)以及正常和微弱动脉搏动情况下进行了评估。使用连接Y形接头适配器的单管袖带不会对监测仪的性能产生不利影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验