Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures.

作者信息

Rooke G A, Bowdle T A

机构信息

Department of Anesthesiology, University of Washington School of Medicine, Seattle.

出版信息

Anesth Analg. 1994 Jan;78(1):150-6. doi: 10.1213/00000539-199401000-00024.

Abstract

Syringe pumps for vasoactive infusions have the advantages of small size and weight, portability, and low cost of the disposable components. However, limited syringe capacity necessitates the use of high drug concentrations, and the accidental delivery of even a small volume of infusate could seriously alter the patient's hemodynamics. To determine the circumstances under which drug delivery might be delayed, or inadvertent boluses could be delivered into the manifold, two brands of commercially available clinical syringe pumps were connected to a stopcock manifold via small-bore tubing and a series of tests were performed. When the syringe pumps were operated at 3 mL/h against a closed stopcock, the pumps' occlusion alarms did not sound for 18-22 min, and when the stopcock subsequently was opened, 0.6-0.9 mL of infusate was delivered as a bolus. Elevating the syringe pump by 120 cm resulted in the delivery of up to 0.5 mL of infusate with the pump turned off. When a syringe pump operating at 6 mL/h was turned off, typically an additional 0.05 mL was delivered during the ensuing 2-3 min. Depending upon the method used to flush the tubing prior to use, delays in drug delivery of 2-3 min occurred at an infusion rate of 3 mL/h. These observations emphasize the need for careful equipment setup and proper use of the manifold stopcocks to avoid unintended drug administration or delay in drug administration.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验