斗篷野照射的准确性:照射体积与每日剂量。

Accuracy in mantle field irradiations: irradiated volume and daily dose.

作者信息

Weltens C, Leunens G, Dutreix A, Cosset J M, Eschwege F, van der Schueren E

机构信息

Department of Radiotherapy, UH Stint Rafaël, Leuven, Belgium.

出版信息

Radiother Oncol. 1993 Oct;29(1):18-26. doi: 10.1016/0167-8140(93)90168-8.

Abstract

Accuracy of patient positioning and dose delivery in mantle field irradiations was investigated on 29 patients treated between August 1990 and December 1991. Patients were treated in two different centers, University Hospital St. Rafaël in Leuven and Institut Gustave Roussy in Villejuif, where different techniques and procedures were used. Measurements were performed on 341 portal films and entrance doses were measured in 518 treatment set-ups. The impact of systematic errors occurring during treatment preparation and day-to-day variations on the accuracy of treatment execution were separately analysed. Daily reproducibility, defined as the deviation from the respective mean measured value for a treatment was demonstrated to be good for both the treated volume and the delivered dose and no difference between the two techniques was shown. Comparing the successive portal films of individual patients (reproducibility of a treatment, once it has started), only small day-to-day variations are found: the SD is 3.4 mm for craniocaudal movements and 2.6 mm for lateral movements. For dose delivery very narrow distributions are obtained with SDs of, respectively, 1.5% and 1.85% for the Leuven and the Villejuif group. This suggests that the position of the patient, which is often thought as the critical point in this complex set-up can be done in a very accurate way, regardless of the position used. To assess the global accuracy of the treatment, the actually treated volume and delivered dose were compared with the planned values. Apart from reproducibility this also takes into account the whole preparatory procedure between planning and the start of the therapy (first session).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

对1990年8月至1991年12月期间接受治疗的29例患者进行了斗篷野照射中患者体位摆放和剂量输送准确性的研究。患者在两个不同的中心接受治疗,分别是鲁汶的圣拉斐尔大学医院和维勒瑞夫的古斯塔夫·鲁西研究所,在这两个中心采用了不同的技术和程序。对341张射野片进行了测量,并在518次治疗设置中测量了入射剂量。分别分析了治疗准备过程中出现的系统误差和日常变化对治疗执行准确性的影响。每日重复性定义为一次治疗相对于各自测量平均值的偏差,结果表明,对于治疗体积和输送剂量而言,重复性都很好,且两种技术之间未显示出差异。比较个体患者连续的射野片(治疗开始后的治疗重复性),仅发现微小的日常变化:头脚方向移动的标准差为3.4毫米,横向移动的标准差为2.6毫米。对于剂量输送,鲁汶组和维勒瑞夫组的标准差分别为1.5%和1.85%,分布非常窄。这表明,通常被认为是这种复杂设置中关键点的患者体位摆放可以非常准确地完成,无论采用何种体位。为了评估治疗的整体准确性,将实际治疗体积和输送剂量与计划值进行了比较。除了重复性之外,这还考虑了从计划到治疗开始(第一疗程)之间的整个准备过程。(摘要截选至250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验