Mock U, Dieckmann K, Molitor A M, Haverkamp U, Pötter R
Department of Radiotherapy and Radiobiology, University Hospital, Vienna, Austria.
Strahlenther Onkol. 1998 Oct;174(10):529-33. doi: 10.1007/BF03038986.
Radiation treatment for malignant lymphoma requires large field irradiation with irregular blocks according to the individual anatomy and tumor configuration. For determination of safety margins (PTV) we quantitatively analysed the accuracy of field and block placement with regard to different anatomical regions.
Forty patients with malignant lymphoma were irradiated using the classical supra-/infradiaphragmatic field arrangements. Treatment was performed with 10-MeV photons and irregularly shaped, large opposing fields. We evaluated the accuracy of field and block placements during the treatment courses by comparing the regularly performed verification-with the simulation films. Deviations were determined with respect to the field edges and the central axis, along the x- and z-axis.
With regard to the field edges, mean deviations of 2.0 mm and 3.4 mm were found along the x- and z-axis. The corresponding standard deviations were 3.4 mm and 5.5 mm, respectively. With regard to the shielding blocks, mean displacement along the x- and z-axis was 2.2 mm and 3.8 mm. In addition, overall standard deviations of 5.7 mm (x-axis) and 7.1 mm (z-axis) were determined. During the course of time an improved accuracy of block placement was notable.
Systematic analysis of port films gives information for a better defining safety margins in external radiotherapy. Evaluation of verification films on a regular basis improves set-up accuracy by reducing displacements.
恶性淋巴瘤的放射治疗需要根据个体解剖结构和肿瘤形态进行大野照射并使用不规则挡块。为了确定安全边界(计划靶区),我们定量分析了不同解剖区域野和挡块放置的准确性。
40例恶性淋巴瘤患者采用经典的膈上/膈下野照射方案。使用10兆电子伏光子和形状不规则的大的对穿野进行治疗。我们通过将定期进行的验证与模拟片进行比较,评估了治疗过程中野和挡块放置的准确性。沿着x轴和z轴,相对于野边缘和中心轴确定偏差。
关于野边缘,在x轴和z轴上的平均偏差分别为2.0毫米和3.4毫米。相应的标准差分别为3.4毫米和5.5毫米。关于屏蔽挡块,在x轴和z轴上的平均位移分别为2.2毫米和3.8毫米。此外,确定了x轴的总体标准差为5.7毫米,z轴为7.1毫米。在治疗过程中,挡块放置的准确性有明显提高。
对射野片进行系统分析可为更好地确定外照射放疗的安全边界提供信息。定期评估验证片可通过减少位移来提高摆位准确性。