Clinical evaluation of the macroduct sweat collection system and conductivity analyzer in the diagnosis of cystic fibrosis.

作者信息

Hammond K B, Turcios N L, Gibson L E

机构信息

Department of Pediatrics, University of Colorado School of Medicine, Denver.

出版信息

J Pediatr. 1994 Feb;124(2):255-60. doi: 10.1016/s0022-3476(94)70314-0.

Abstract

The purposes of this study were to compare sweat tests used in diagnosing cystic fibrosis (CF), as performed with the Macroduct collection system, with those utilizing the more laborious quantitative pilocarpine iontophoresis test (QPIT), and to ascertain the efficacy of the Sweat-Chek conductivity analyzer in eliminating some possibly unnecessary chloride analyses. A Macroduct sweat test was performed on one arm and a QPIT on the other on 1090 patients, 93 of whom had CF. Of these, 514 patients (43 with CF) also had a conductivity determination on the Macroduct sweat sample. All subjects were referred to the laboratory of one of us (K.B.H.) for sweat testing. Of the QPIT samples, 0.7% were inadequate, as were 6.1% of those from the Macroduct system. When sodium and chloride concentrations from the two tests were compared, the standard errors of the estimate were 3.90 and 3.85, respectively. Agreement within 8 mEq/L could then be expected with 95% confidence limits. With use of the Sweat-Chek analyzer, no patient with CF was found to have a conductivity of less than 90 mmol/L, whereas 430 (91%) of the non-CF subjects had a conductivity of less than 50 mmol/L. None of those 430 subjects had a sweat chloride value > 32 mmol/L. We conclude that the Macroduct collection system provides results equally as satisfactory as those provided by the QPIT and that the Sweat-Chek analyzer frequently eliminates the necessity of measuring chloride concentrations.

摘要

本研究的目的是比较使用Macroduct收集系统进行的用于诊断囊性纤维化(CF)的汗液测试和使用更繁琐的定量毛果芸香碱离子导入测试(QPIT)的汗液测试,并确定Sweat-Chek电导率分析仪在消除一些可能不必要的氯化物分析方面的功效。对1090例患者的一侧手臂进行Macroduct汗液测试,另一侧手臂进行QPIT测试,其中93例患有CF。其中,514例患者(43例患有CF)还对Macroduct汗液样本进行了电导率测定。所有受试者均被转至我们其中一人(K.B.H.)的实验室进行汗液测试。QPIT样本中有0.7%不合格,Macroduct系统的样本中有6.1%不合格。比较两种测试的钠和氯化物浓度时,估计的标准误差分别为3.90和3.85。然后可以预期在8 mEq/L范围内的一致性,置信限为95%。使用Sweat-Chek分析仪时,未发现CF患者的电导率低于90 mmol/L,而非CF受试者中有430例(91%)的电导率低于50 mmol/L。这430名受试者中没有一人的汗液氯化物值>32 mmol/L。我们得出结论,Macroduct收集系统提供的结果与QPIT提供的结果同样令人满意,并且Sweat-Chek分析仪经常消除测量氯化物浓度的必要性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验