Nelson S, Pritt A, Marlar R A
Special Coagulation Laboratory, Denver Veterans Affairs Medical Center, CO.
Arch Pathol Lab Med. 1994 Feb;118(2):175-6.
We undertook to confirm and extend the previous work on the use of microcentrifugation (2 minutes at 11,000 g) to prepare platelet-poor plasma for assay of the prothrombin time, partial thromboplastin time, fibrinogen level, D-dimer, antithrombin III, and dilute Russell viper venom time. We compared results of routinely submitted blood samples by both high-speed and routine (15 minutes at 1800g) centrifugation. We found no significant differences in assay results and concluded that the high-speed technique is a reliable and useful option for minimizing turnaround times for these coagulation assays.