Comparison of the Keeler Pulsair 2000 non-contact tonometer with Goldmann applanation.

作者信息

Moseley M J, Thompson J R, Deutsch J, Misson G P, Naylor G, Tan-Yee A, Taylor R H, Fielder A R

机构信息

Department of Ophthalmology, University of Birmingham, UK.

出版信息

Eye (Lond). 1993;7 ( Pt 1):127-30. doi: 10.1038/eye.1993.26.

Abstract

The Pulsair 2000 non-contact tonometer (Keeler Ltd, UK) is compared with the Goldmann applanation tonometer. Data from 80 eyes were acquired by four experienced observers. A linear regression analysis showed the relationship between the instruments to be: Pulsair = 0.66 + 0.95 Goldmann. Individual components of variation were analysed by analysis of variance which indicated a significant variation in the slope of the regression equation due to observers (p = 0.02) but not to the order in which topical anaesthesia was administered. Differences between two Pulsair instruments were of marginal significance (p = 0.07). The intercept of the regression equation was unaffected by any of the components of variation. Seventy-nine per cent of averaged intraocular pressure measurements obtained with the Pulsair 2000 fell on or within +/- 3 mmHg of those measured with the Goldmann tonometer. It is concluded that the Pulsair 2000 can provide clinically useful measurements of intraocular pressure.