[Comparative study of pneumotonometer and Goldmann tonometer for screening high intraocular pressure in primary care].

作者信息

Jara Peñacoba M, López Traynor A, Duce Tello S, Navas Serrano V, González Sanz M, Toledano Fernández N

机构信息

Centro de Salud Luengo Rodríguez, Móstoles, Madrid.

出版信息

Aten Primaria. 2000 Apr 30;25(7):493-6. doi: 10.1016/s0212-6567(00)78550-7.

Abstract

OBJECTIVE

Evaluation of reliability of measurements of IOP obtained with air-puff noncontact tonometer respect the obtained with the conventional Goldmann tonometer. As well, analysis of the possibility of a higher difference between the measurements by both methods respect the fact of being myope or hypermetrope.

DESIGN

Comparative study of two measurement methods.

SETTING

General ophthalmology clinic and primary care clinic of our sanitary area.

PATIENTS

Aleatory sample among the patients who went to the general ophthalmology clinic for any cause from the first of May to the thirtieth of June of 1996. The patients with predisposing processes for glaucoma were excluded, as well as those who suffered ocular surgery or those who took drugs which influenced the IOP.

INTERVENTIONS

The ophthalmologist made three measurements of IOP in each eye and the sight of all the patients were tested. The arithmetic average was made among every three measurements. The physician of primary care made three measurements of IOP in each eye with the air-puff noncontact tonometer, and the arithmetic average was made.

RESULTS

81 patients were included, from whom 7 presented high IOP. The sensibility of the air-puff noncontact tonometer, compared with Goldmann tonometer, was 86% (95% CI, 18.20-99.63%), and the specificity 84% (95% CI, 89.66-78.08%). The air-puff noncontact tonometer obtained measurements between 1.116 and 2.008 mmHg higher than the Goldman tonometer. This difference, worthless from the clinic point of view, didn't find a relationship with the fact of being myope or hypermetrope. It was found a positive lineal association between the measurements made by both methods, with a correlation coefficient of 0.8086 (p < 0.001, 95% CI, 0.7476-0.8560).

CONCLUSIONS

The results obtained are similar to the ones of other published series. A tendency of higher measurements with the air-puff noncontact tonometer is observed. It hasn't been observed a higher difference between both methods in myopes. The air-puff tonometer is a valid reliable technique to be used in primary care, it is easy to use, it doesn't transmit infectious illnesses, and it isn't necessary to use anaesthetic or staining eyedrops.

摘要

相似文献

4
Accuracy of the Tonosafe disposable tonometer head compared to the Goldmann tonometer alone.
Clin Exp Ophthalmol. 2004 Aug;32(4):364-7. doi: 10.1111/j.1442-9071.2004.00838.x.
5
[Usefulness of aplanation tonometry in primary care].
Aten Primaria. 1996 Sep 30;18(5):217-20.
6
A new tonometer--the Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation tonometers.
Invest Ophthalmol Vis Sci. 2013 Jan 23;54(1):659-65. doi: 10.1167/iovs.12-10984.
7
Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma.
Clin Exp Ophthalmol. 2004 Dec;32(6):584-9. doi: 10.1111/j.1442-9071.2004.00907.x.
8
Corneal biomechanical properties: precision and influence on tonometry.
Cont Lens Anterior Eye. 2014 Jun;37(3):124-31. doi: 10.1016/j.clae.2013.09.006. Epub 2013 Oct 11.
9
[Clinical evaluation of the Pascal dynamic contour tonometer].
J Fr Ophtalmol. 2007 Mar;30(3):260-70. doi: 10.1016/s0181-5512(07)89588-x.

本文引用的文献

1
Medical management of glaucoma.
N Engl J Med. 1998 Oct 29;339(18):1298-307. doi: 10.1056/NEJM199810293391808.
3
Comparison of the Keeler Pulsair 2000 non-contact tonometer with Goldmann applanation.
Eye (Lond). 1993;7 ( Pt 1):127-30. doi: 10.1038/eye.1993.26.
4
Comparative study of three non-contact tonometers and the Goldmann tonometer.
Ophthalmologica. 1994;208(3):115-8. doi: 10.1159/000310466.
5
Reproducibility with the Keeler Pulsair 2000 non-contact tonometer.
Br J Ophthalmol. 1995 Jun;79(6):554-7. doi: 10.1136/bjo.79.6.554.
6
Comparison of a new non-contact tonometer with Goldmann applanation.
Eye (Lond). 1989;3 ( Pt 3):332-7. doi: 10.1038/eye.1989.48.
9
Clinical evaluation of two non-contact tonometers.
Doc Ophthalmol. 1991;78(3-4):259-63. doi: 10.1007/BF00165689.
10

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验