J Appl Physiol Respir Environ Exerc Physiol. 1977 Jan;42(1):120-3. doi: 10.1152/jappl.1977.42.1.120.
Equations for the calculation of VO2, VCO2, and evaporative water loss were derived for use of a paramagnetic gas analyzer with a flow-through respirometry system and an open or closed mask. The magnitude of errors involved in the determination of VO2 with no CO2 absorbent are small (+/-3%) but may be greater if no H2O absorbent is used, unless an appropriate correction is made for evaporative water loss. Data collected for hummingbirds (Selasphorus sasin) and monitor lizards (Varanus gouldii) indicate the validity of the technics described for the measurement of RQ and evaporative water loss and demonstrate the use of paramagnetic gas analyzers in monitoring respiratory patterns.
推导了用于通过顺磁气体分析仪与流通式呼吸测定系统以及开放式或封闭式面罩来计算VO₂、VCO₂和蒸发失水量的方程式。在没有二氧化碳吸收剂的情况下测定VO₂时所涉及的误差幅度较小(±3%),但如果不使用水吸收剂,误差可能会更大,除非对蒸发失水量进行适当校正。对蜂鸟(红喉北蜂鸟)和巨蜥(古氏巨蜥)收集的数据表明了所描述的用于测量呼吸商和蒸发失水量技术的有效性,并证明了顺磁气体分析仪在监测呼吸模式中的应用。