Film density calibration for computed radiography systems: is the standard three-point procedure accurate?

作者信息

Huda W, Jing Z, Hoyle B A

机构信息

Department of Radiology, College of Medicine, J. Hillis Miller Health Center, University of Florida, Gainesville 32610-0374.

出版信息

Radiology. 1993 Sep;188(3):875-7. doi: 10.1148/radiology.188.3.8351366.

Abstract

With computed radiography (CR) systems, laser printer-film processor combinations are normally checked daily by using test film three-point density measurements. In this study, the authors show that the recommended three-point quality assurance procedure can show satisfactory results even though the CR film output behaves in an anomalous manner. This problem can be corrected by using a 16-point calibration procedure. Regular 16-point calibrations should be performed to ensure satisfactory CR system performance.

摘要

对于计算机X线摄影(CR)系统,通常每天使用测试胶片进行三点密度测量来检查激光打印机 - 胶片处理器组合。在本研究中,作者表明,即使CR胶片输出表现异常,推荐的三点质量保证程序仍可显示出令人满意的结果。通过使用16点校准程序可以纠正此问题。应定期进行16点校准,以确保CR系统性能令人满意。