Phantom pacemaker programming.

作者信息

Fieldman A, Dobrow R J

出版信息

Pacing Clin Electrophysiol. 1978 Apr;1(2):166-71. doi: 10.1111/j.1540-8159.1978.tb03460.x.

Abstract

Spurious or "phantom" programming of programmable pacemakers may occur as the result of two phenomena. One is change in rate and/or output of the generator from an anomalous source (dysprogramming); the other is a generator response different from that intended and set on the conventional programmer (misprogramming). In either case, the result is a nuisance, at best. Conceivably, at worst, an irreversible subthreshold output could be induced. Despite wide usage of such devices, neither event has been reported. In the routine follow-up of 95 implanted Cordis Omni-Stanicor generators, we have observed three instances of dysprogramming upon application of a magnet commonly used to induce the asynchronous mode. Subsequent in vitro studies revealed magnetically induced programming reed switch vibration in 9 of 6,680 units, which could account for the in vivo observations. We also describe many instances of misprogramming, attributable, in most cases, to a faulty programmer emission count. In this series, none of the anomalies was clinically detrimental. However, anyone involved in pacemaker follow-up should be alerted to the existence of these potentially hazardous phenomena.

摘要

可编程起搏器可能会出现假性或“幻影”编程,这是由两种现象导致的。一种是发生器的速率和/或输出因异常源而改变(程序紊乱);另一种是发生器的反应与传统编程器上设定的预期反应不同(编程错误)。无论哪种情况,结果充其量只是麻烦事。可以想象,在最坏的情况下,可能会诱发不可逆的阈下输出。尽管此类装置被广泛使用,但尚未有这两种情况的报道。在对95个植入的科迪斯全向型起搏器发生器进行的常规随访中,我们观察到在应用常用于诱发异步模式的磁铁时出现了3例程序紊乱情况。随后的体外研究显示,在6680个装置中有9个出现了由磁性诱发的编程簧片开关振动,这可以解释体内观察到的情况。我们还描述了许多编程错误的实例,在大多数情况下,这归因于编程器发射计数错误。在这个系列中,没有一个异常情况对临床有不利影响。然而,任何参与起搏器随访的人员都应警惕这些潜在危险现象的存在。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验