[心脏直视手术后患者有创与无创血压测量的比较]

[Comparison of invasive and non-invasive measuring of blood pressure in patients following open heart surgery].

作者信息

Lupkovics G, Rudas L

机构信息

Szívsebészeti Onálló Osztály, Szent-Györgyi Albert Orvostudományi Egyetem, Szeged.

出版信息

Orv Hetil. 1993 Sep 12;134(37):2033-5.

PMID:8414448
Abstract

Accuracy of blood pressure measurements with a new noninvasive ambulatory blood pressure monitor (Meditech KFT Budapest) was studied in 12 postoperative patients following open heart surgery. We compared 532 systolic and diastolic blood pressure recordings taken simultaneously both invasively and noninvasively. The values obtained noninvasively were subtracted from the corresponding values obtained invasively, and the differences were assessed according the British Hypertension Society recommendations. Values obtained with the two methods demonstrated good correlation (R: 0.85; p < 0.0001). 60 per cent of the differences in diastolic and 76 per cent of the differences in systolic blood pressure recordings fell into the range of +/- 5 mmHg. Thus the accuracy of diastolic blood pressure measurements was classified as "A" (characteristic fo the best equipment) and those of the systolic recordings was classified as "B" (characteristic of good equipment). Based on our findings the new blood pressure monitor provides accurate recordings and its use is highly recommended in the everyday practice.

摘要

我们对12例心脏直视手术后的患者,研究了一种新型无创动态血压监测仪(匈牙利布达佩斯Meditech KFT公司生产)测量血压的准确性。我们对532次同时采用有创和无创方式记录的收缩压和舒张压进行了比较。将无创测量值从相应的有创测量值中减去,并根据英国高血压学会的建议评估差值。两种方法获得的值显示出良好的相关性(R:0.85;p < 0.0001)。舒张压记录差异的60%和收缩压记录差异的76%落在±5 mmHg范围内。因此,舒张压测量的准确性被归类为“A”(最佳设备的特征),收缩压记录的准确性被归类为“B”(良好设备的特征)。基于我们的研究结果,这种新型血压监测仪能提供准确的记录,强烈建议在日常实践中使用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验