We present a new computer-controlled, electromechanical system for performing simultaneous extension, inflation, and torsion experiments on cylindrical segments of natural and artificial blood vessels. Specimens are tested while immersed in a temperature-controlled, oxygenated, physiologic solution. Deformations are measured within a central region of the specimen using noncontacting video methods. The associated axial loads, luminal pressures, and torques are measured with standard transducers. Data are collected and stored on-line, and are used in the feedback control of experimental protocols, which are prescribed using custom interactive software. Finally, we present illustrative data obtained from canine aortas and common carotid arteries.

摘要

我们展示了一种新型的计算机控制的机电系统,用于对天然和人造血管的圆柱形节段同时进行拉伸、膨胀和扭转实验。样本在浸没于温度受控、充氧的生理溶液中时接受测试。使用非接触式视频方法在样本的中心区域内测量变形。相关的轴向载荷、管腔内压力和扭矩通过标准传感器进行测量。数据在线收集和存储,并用于实验方案的反馈控制,实验方案是使用定制的交互式软件规定的。最后,我们展示了从犬主动脉和颈总动脉获得的说明性数据。