Monsein L H, Kelsey C A, Williams W L, Olson N J
Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins University School of Medicine, Baltimore, Maryland.
Cardiovasc Intervent Radiol. 1993 Jan-Feb;16(1):11-3. doi: 10.1007/BF02603030.
Aspiration biopsies of fresh disease-free bovine hepatic liver were performed with three commercially available, small-gauge biopsy needles. A syringe was used to create a negative pressure when using Chiba and Sure-cut needles. A syringe was not used when using PercuCut needles which create an internal negative pressure equivalent to a 1.5 ml syringe when the stylet is partially withdrawn. The PercuCut and Chiba needle yielded a significantly larger specimen size than the Sure-Cut needle. There was no significant difference in sample size between the PercuCut and Chiba needles. All samples were of diagnostic quality.