Kreula J, Virkkunen P, Bondestam S
Department of Diagnostic Radiology, Helsinki University Central Hospital, Finland.
Invest Radiol. 1990 Nov;25(11):1175-81. doi: 10.1097/00004424-199011000-00005.
High incidences of unsatisfactory specimens in fine-needle aspiration biopsy for cytology have been reported previously. Inappropriate aspiration techniques may account partly for this. In this study, the effect of suction on specimen size was studied in vitro using an automatic sampler. Bovine liver, testis, and pig's kidney were used as target tissues and biopsies were performed with 18-, 20- and 22-gauge needles. Specimen weight increased rapidly and linearly with suction force (P less than .001), regardless of the target tissue or needle diameter, when standard needle movement was applied. In the absence of needle movement, the suction force, rate of increase of suction, and duration of suction had no effect on the weight of the specimens. Needle movement alone, without suction, also produced meager specimens. Suction alone does not seem to tear fragments off target tissues. Nevertheless, in combination with needle movement, suction force plays an important role in the determination of specimen size.
先前已有报道称,细针穿刺活检用于细胞学检查时,不满意标本的发生率较高。抽吸技术不当可能是部分原因。在本研究中,使用自动采样器在体外研究了吸力对标本大小的影响。将牛肝、睾丸和猪肾用作靶组织,并用18号、20号和22号针进行活检。当采用标准针移动时,无论靶组织或针直径如何,标本重量均随吸力迅速且呈线性增加(P<0.001)。在没有针移动的情况下,吸力、吸力增加速率和抽吸持续时间对标本重量没有影响。仅针移动而无吸力也只能产生少量标本。仅靠吸力似乎不会从靶组织上撕下碎片。然而,与针移动相结合时,吸力在标本大小的确定中起着重要作用。