Evaluation of a new ambulatory spirometer for measuring forced expiratory volume in one second and peak expiratory flow rate.

作者信息

Johns D P, Abramson M, Bowes G

机构信息

Department of Respiratory Medicine, Alfred Hospital, Prahran, Victoria, Australia.

出版信息

Am Rev Respir Dis. 1993 May;147(5):1245-50. doi: 10.1164/ajrccm/147.5.1245.

Abstract

A new pocket-sized ambulatory spirometer (meter) that measures FEV1 and PEFR was evaluated. The meter determines flow rate (V) from the differential pressure (P1-P2) across a sharp-edged orifice (V alpha square root of[P1-P2]). Ten meters were evaluated for accuracy using a computerized syringe to deliver the 24 ATS-recommended waveforms and nine scaled versions of a single waveform (PW#24). Within-meter reproducibility of FEV1 and PEFR was evaluated in two meters by passing five versions of waveform PW#24 through each meter 10 times. Reliability in the ambulatory setting was assessed in six meters on several occasions over a 10-week period using five versions of waveform PW#24. Results show that the 10 meters conform to the ATS accuracy specifications for PEFR with one or less errors and marginally outside these limits for FEV1 with four errors. For the nine versions of PW#24, the 95% confidence intervals indicate that the meter is accurate to within +/- 5.5% or +/- 15 L/min for PEFR and +/- 3.5% or +/- 0.12 L for FEV1. The mean within-meter coefficient of variation was 1.24% for FEV1 and 0.35% for PEFR. There was no significant change in meter accuracy or performance over the 10-wk reliability study. We conclude that the meter is suitable for use as an ambulatory spirometer for measuring FEV1 and PEFR.

摘要

对一种新型袖珍式便携式肺量计(仪器)进行了评估,该仪器可测量第一秒用力呼气量(FEV1)和呼气峰值流速(PEFR)。该仪器通过测量跨过锐边孔口的压差(P1 - P2)来确定流速(V)(V与[P1 - P2]的平方根成正比)。使用计算机控制的注射器输送24种美国胸科学会(ATS)推荐的波形以及单个波形(PW#24)的九个缩放版本,对十台仪器的准确性进行了评估。通过让五个版本的波形PW#24分别通过两台仪器10次,评估了两台仪器内FEV1和PEFR的重复性。在10周的时间里,多次使用五个版本的波形PW#24,对六台仪器在动态环境中的可靠性进行了评估。结果表明,这十台仪器在PEFR方面符合ATS准确性规范,误差为一个或更少,而在FEV1方面略超出这些限制,有四个误差。对于PW#24的九个版本,95%置信区间表明该仪器对于PEFR的测量精度在±5.5%或±15升/分钟以内,对于FEV1的测量精度在±3.5%或±0.12升以内。仪器内FEV1的平均变异系数为1.24%,PEFR为0.35%。在为期10周的可靠性研究中,仪器的准确性和性能没有显著变化。我们得出结论,该仪器适合用作测量FEV1和PEFR的便携式肺量计。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验