Suppr超能文献

一种新型便携式电话传输肺量计的评估

Assessment of a new transtelephonic portable spirometer.

作者信息

Abboud S, Bruderman I

机构信息

Biomedical Engineering Department, Tel Aviv University, Israel.

出版信息

Thorax. 1996 Apr;51(4):407-10. doi: 10.1136/thx.51.4.407.

Abstract

BACKGROUND

A new portable spirometer, the Spirophone, has been developed that records a subject's blow and can then transmit all the data by telephone to a receiving centre for analysis and comment. Tests of this device were undertaken to determine its accuracy and reliability.

METHODS

The performance of the Spirophone was tested using computer generated wave forms, by delivering blows from calibrated syringes at different flows, and by comparing subjects' blows with those recorded with a commercial spirometer.

RESULTS

Using computer generated wave forms all lung function indices were accurate to within 1% and blows delivered from calibrating syringes were accurate to within 5%. When subjects performed repeated forced vital capacity (FVC) manoeuvres there were no significant differences between lung function indices recorded with the Spirophone and with a commercial spirometer. With the Spirophone and commercial spirometer in series the FVC and forced expiratory volume in one second (FEV1) were within 5% of each other in nine out of 10 healthy subjects.

CONCLUSION

The Spirophone recorded maximal forced expiratory manoeuvres with acceptable accuracy, reliability, and reproducibility, and this system offers the ability to monitor a patient's lung function at a centre remote from the patient.

摘要

背景

已研发出一种新型便携式肺活量计——呼吸听诊式肺活量计(Spirophone),它可记录受试者的呼气,并能通过电话将所有数据传输至接收中心进行分析和评论。对该设备进行了测试以确定其准确性和可靠性。

方法

使用计算机生成的波形、通过校准注射器以不同流速进行呼气以及将受试者的呼气与商用肺活量计记录的呼气进行比较,对呼吸听诊式肺活量计的性能进行了测试。

结果

使用计算机生成的波形时,所有肺功能指标的误差在1%以内,校准注射器呼出的气量误差在5%以内。当受试者进行重复的用力肺活量(FVC)操作时,呼吸听诊式肺活量计和商用肺活量计记录的肺功能指标之间无显著差异。在10名健康受试者中,有9名受试者使用串联的呼吸听诊式肺活量计和商用肺活量计时,其FVC和一秒用力呼气量(FEV1)相差在5%以内。

结论

呼吸听诊式肺活量计记录最大用力呼气操作时具有可接受的准确性、可靠性和可重复性,并且该系统能够在远离患者的中心监测患者的肺功能。

相似文献

本文引用的文献

4
Evaluating commercially available spirometers.评估市售肺活量计。
Am Rev Respir Dis. 1980 Jan;121(1):73-82. doi: 10.1164/arrd.1980.121.1.73.
5
The effect of temperature on recording spirograms.温度对记录呼吸描记图的影响。
Am Rev Respir Dis. 1983 Nov;128(5):894-8. doi: 10.1164/arrd.1983.128.5.894.
6
Evaluation of "electronic" spirometers.“电子”肺活量计的评估
N Engl J Med. 1973 Dec 13;289(24):1283-8. doi: 10.1056/NEJM197312132892406.
7
Heated Fleisch pneumotachometer: a calibration procedure.热式弗莱施呼吸速度计:校准程序。
J Appl Physiol. 1973 Jan;34(1):117-21. doi: 10.1152/jappl.1973.34.1.117.
8
Linearity and temperature control of the Fleisch pneumotachograph.弗莱施呼吸流速计的线性度和温度控制
J Appl Physiol (1985). 1986 Feb;60(2):710-5. doi: 10.1152/jappl.1986.60.2.710.
9
Evaluation of the turbine pocket spirometer.涡轮式袖珍肺活量计的评估
Thorax. 1987 Sep;42(9):689-93. doi: 10.1136/thx.42.9.689.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验