Stevens M, Prentice M B, Swann R A, Mitchell C J, Donkin A
Public Health Laboratory, Leicester Royal Infirmary, UK.
J Appl Bacteriol. 1993 Apr;74(4):417-20. doi: 10.1111/j.1365-2672.1993.tb05148.x.
The Signal blood culture system was compared with the Bactec NR-660. A total of 1617 blood culture sets yielded 143 (8.8%) significant isolates; 113 (79.0%) were from positive bottles in both the Bactec and Signal systems. Twelve organisms (8.4%) were detected and isolated from the Signal system only and another 18 (12.6%) from the Bactec system only. Of these 18, five were Signal-positive but the organism was not recovered and four organisms were isolated from negative Signal bottles on terminal subculture. The time taken to detection for each system was similar; the Signal system detected 68% and the Bactec 63% of significant positives within 24 h. At 48 h Bactec detected 91% and the Signal 85%. A significantly-reduced number of bottles which gave a positive signal but were negative by microscopical and cultural methods was found, compared with previous reports. The 1 h incubation period prior to the insertion of the Signal growth indicator device was considered to be the cause of this reduction in the proportion of false positives. Fifty-five percent (42/77) of the Bactec false positives were due to delta growth value. This is when there is an increase in the growth index of > or = 15 without the positive threshold level of 30 being attained. This occurred in the anaerobic bottle on day 2 with 42 bottles. Another 40% (31/77) of the false positives had a growth value between the positive threshold of 30 and a value of 35. Eighty (4.9%) of Bactec and 65 (4.0%) of Signal sets yielded clinically non-significant isolates.(ABSTRACT TRUNCATED AT 250 WORDS)
将信号血培养系统与Bactec NR - 660进行了比较。总共1617套血培养标本培养出143株(8.8%)有意义的分离株;其中113株(79.0%)在Bactec和信号系统的阳性瓶中均被检出。仅在信号系统中检测到并分离出12种微生物(8.4%),仅在Bactec系统中检测到并分离出另外18种(12.6%)。在这18种中,有5种在信号系统中呈阳性但未培养出微生物,还有4种微生物是在信号瓶终末传代培养时从阴性瓶中分离出来的。每个系统检测所需时间相似;信号系统在24小时内检测到68%的有意义阳性结果,Bactec系统检测到63%。在48小时时,Bactec系统检测到91%,信号系统检测到85%。与之前的报告相比,发现发出阳性信号但经显微镜和培养方法检测为阴性的瓶子数量显著减少。信号生长指示装置插入前1小时的孵育期被认为是假阳性比例降低的原因。Bactec系统55%(42/77)的假阳性是由于δ生长值。这是指生长指数增加≥15但未达到阳性阈值水平30的情况。在第2天,42个厌氧瓶出现这种情况。另外40%(31/77)的假阳性生长值在阳性阈值30和35之间。Bactec系统80套(4.9%)和信号系统65套(4.0%)培养出临床无意义的分离株。(摘要截短于250字)