Ma C M, Rogers D W, Shortt K R, Ross C K, Nahum A E, Bielajew A F
Physics Department, Royal Marsden Hospital, Sutton, Surrey, United Kingdom.
Med Phys. 1993 Mar-Apr;20(2 Pt 1):283-92. doi: 10.1118/1.597128.
For megavoltage radiotherapy photon beams, EGS4 Monte Carlo calculations show, and experimental measurements confirm with an accuracy of 0.2%, that glass or quartz-walled vials used in Fricke dosimetry increase the dose in the Fricke solution. This is mainly caused by increased electron scattering from the glass which increases the dose to the Fricke solution. The dose perturbation is shown to vary from nothing in a 60Co beam up to 2% in a 24-MV beam. For plastic vials of similar shapes, calculations demonstrate that the effect is in the opposite direction and even at high energies it is much less (0.2% to 0.5%).
对于兆伏级放疗光子束,EGS4蒙特卡罗计算表明,实验测量也以0.2%的精度证实,弗里克剂量测定法中使用的玻璃或石英壁小瓶会增加弗里克溶液中的剂量。这主要是由于玻璃电子散射增加,从而增加了弗里克溶液的剂量。剂量扰动在60Co束中为零,在24兆伏束中高达2%。对于形状相似的塑料小瓶,计算表明这种效应方向相反,即使在高能情况下也小得多(0.2%至0.5%)。