节段间关节动力学与等速测力计测量结果的比较。

A comparison of intersegmental joint dynamics to isokinetic dynamometer measurements.

作者信息

Kaufman K R, An K N, Chao E Y

机构信息

Motion Analysis Laboratory, Children's Hospital, San Diego, CA 92123, USA.

出版信息

J Biomech. 1995 Oct;28(10):1243-56. doi: 10.1016/0021-9290(94)00176-5.

Abstract

This study quantified the errors associated with measurements obtained using an isokinetic dynamometer for knee evaluation at 60 degrees and 180 degrees s-1. Five normal subjects were tested. A triaxial electrogoniometer was used to measure the kinematics. A custom-designed load cell was used to measure the three orthogonal components of the reaction force applied by the isokinetic dynamometer to the leg. The inverse dynamics problem was solved to obtain the intersegmental dynamics at the knee joint. These data were compared to those obtained from the isokinetic dynamometer. The results indicate that the knee motion recorded by the isokinetic dynamometer was in error by 7-9 degrees during flexion and 10-13 degrees during extension. The isokinetic dynamometer recorded knee flexions which were less than actual during the flexion portion of the exercise and greater than actual during extension. The duration of constant velocity was limited based on the speed selected. The knee moment was in error by 5 Nm during flexion and 25-29 Nm during extension. The use of a gravity correction factor reduced the knee moment error during extension, but increased it during flexion.

摘要

本研究量化了使用等速测力计在60度/秒和180度/秒时进行膝关节评估所获得测量值的相关误差。对五名正常受试者进行了测试。使用三轴电子测角仪测量运动学。使用定制设计的测力传感器测量等速测力计施加于腿部的反作用力的三个正交分量。通过求解逆动力学问题来获得膝关节处的节段间动力学。将这些数据与从等速测力计获得的数据进行比较。结果表明,等速测力计记录的膝关节运动在屈曲时误差为7 - 9度,伸展时误差为10 - 13度。等速测力计记录的膝关节屈曲在运动的屈曲部分小于实际值,在伸展部分大于实际值。基于所选速度,等速阶段的持续时间受到限制。膝关节力矩在屈曲时误差为5牛米,伸展时误差为25 - 29牛米。使用重力校正因子减少了伸展时的膝关节力矩误差,但增加了屈曲时的误差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验