Evaluation of a semi-automated platelet-counting system.

作者信息

Rowan R M, Fraser C, Gray J H, McDonald G A

出版信息

J Clin Pathol. 1977 Apr;30(4):361-6. doi: 10.1136/jcp.30.4.361.

Abstract

Coulter Electronics Ltd have produced a semi-automated platelet-counting system. Platelet-rich plasma may be obtained either by tube sedimentation or by means of the Thrombo-fuge, the latter being an instrument designed to produce accelerated sedimentation. The instrument is linear over the entire range of platelet counts, and machine reproducibility is good. Comparison of machine-rated with visual counts satisfied statistical evaluation. The technique can be handled by one operator and platelet counts can be achieved at the rate of 30 per hour by both methods although individual counts on the Thrombo-fuge may be obtained in approximately one-quarter of the time required for tube sedimentation. The throughput using the Thrombo-fuge could certainly be doubled were two sample plates supplied. Few problems were encountered during the evaluation and most could be avoided by meticulous technique. Visual counts must be performed when the sample haematocrit is greater than 50%-Discrepant counts have been obtained in patients with white cell counts exceeding 50 X 10(9)/1 and in patients with giant platelets. ESR elevation for any reason does not lead to serious discrepancy in results. The incidence of platelet clumping due to the presence of platelet agglutinins and of microclot formation due to inadequate mixing is probably much higher than is commonly thought, and certainly peripheral blood film scrutiny should never be omitted in patients with low counts. Careful examination of peripheral blood films must be combined with instrument counting for some time lest further causes of discrepant counting emerge.

摘要

相似文献

1
Evaluation of a semi-automated platelet-counting system.
J Clin Pathol. 1977 Apr;30(4):361-6. doi: 10.1136/jcp.30.4.361.
2
Automated platelet counting--a re-evaluation of the sedimentation method.
Clin Lab Haematol. 1981;3(3):215-22. doi: 10.1111/j.1365-2257.1981.tb01335.x.
3
Electronic platelet counts with the Coulter counter. Reassessment of a correction factor.
Am J Clin Pathol. 1976 Sep;66(3):506-11. doi: 10.1093/ajcp/66.3.506.
5
An evaluation of the Celloscope 401 electronic blood cell counter.
J Clin Pathol. 1972 Jun;25(6):539-42. doi: 10.1136/jcp.25.6.539.
6
Evaluation of an automatic platelet counting system utilizing whole blood.
J Clin Pathol. 1972 Mar;25(3):218-26. doi: 10.1136/jcp.25.3.218.
8
10
A side-by-side evaluation of four platelet-counting instruments.
Am J Clin Pathol. 1980 Aug;74(2):119-34. doi: 10.1093/ajcp/74.2.119.

引用本文的文献

1
A rapid and accurate differential centrifugation method for platelet counts.
J Clin Pathol. 1982 Jan;35(1):114-6. doi: 10.1136/jcp.35.1.114.

本文引用的文献

1
The reproducibility and constancy of the platelet count.
Am J Clin Pathol. 1953 Jan;23(1):15-26. doi: 10.1093/ajcp/23.1.15.
2
Platelet counts with the Coulter counter.
Am J Clin Pathol. 1965 Dec;44(6):678-88. doi: 10.1093/ajcp/44.6.678.
3
Evaluation of an automatic platelet counting system utilizing whole blood.
J Clin Pathol. 1972 Mar;25(3):218-26. doi: 10.1136/jcp.25.3.218.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验