Piérard G E, Nikkels-Tassoudji N, Piérard-Franchimont C
Laboratory of Dermometrology, University of Liège, Belgium.
Dermatology. 1995;191(1):9-15. doi: 10.1159/000246472.
New advances in bioengineering have provided commercially available devices for measuring the mechanical properties of skin in vivo. Reproducibility of data and methodological approaches have not yet been thoroughly studied.
To study the reproducibility and the influence of the area of the test site on the values of biomechanical variables yielded in a normal population.
A 500-mbar suction was transmitted to the skin through Cutometer probes equipped with a 2- or 8-mm opening.
The best reproducibility was obtained for the maximum distension of skin and for the biological elasticity. The values of the standard biomechanical ratios were almost the same for both probes. Linear correlations were found between parameters of elasticity.
The Cutometer is a reliable device. A high degree of correlation exists between biomechanical variables related to elasticity. Under a suction of 500 mbar, both the 2- and 8-mm probes give results which correlate statistically in a large population. However, the two probes do not measure precisely the same aspect of skin mechanics. The 2-mm probe measures the capacity of superficial skin folding, and the 8-mm probe explores the biomechanical properties of the dermis itself and of its slipping mobility on the hypodermis. The biomechanical parameters which appear the most clinically relevant for a normal-looking skin are the maximum distension and the biological elasticity.
生物工程学的新进展已提供了可用于在体内测量皮肤力学性能的商用设备。数据和方法的可重复性尚未得到充分研究。
研究正常人群中测试部位面积对生物力学变量值的可重复性及影响。
通过配备2毫米或8毫米开口的皮肤弹性测定仪探头向皮肤施加500毫巴的吸力。
皮肤最大扩张度和生物弹性的可重复性最佳。两种探头的标准生物力学比值几乎相同。弹性参数之间存在线性相关性。
皮肤弹性测定仪是一种可靠的设备。与弹性相关的生物力学变量之间存在高度相关性。在500毫巴的吸力下,2毫米和8毫米探头在大量人群中得出的结果在统计学上具有相关性。然而,两种探头并未精确测量皮肤力学的同一方面。2毫米探头测量的是表皮折叠能力,8毫米探头探究的是真皮本身及其在皮下组织上的滑动移动性的生物力学特性。对于外观正常的皮肤,临床上最相关的生物力学参数是最大扩张度和生物弹性。