Cook R T, Moglia B A
Center for Emergency Medical Services, Milton S. Hershey Medical Center, Hershey, Pennsylvania, USA.
J Pediatr. 1996 Apr;128(4):568-70. doi: 10.1016/s0022-3476(96)70373-6.
We evaluated the Beck Airway Air Flow Monitor (BAAM), through which airflow makes a whistling sound, for its safety and efficacy in confirming that an endotracheal tube is in the trachea. We studied 46 neonates ranging in weight from 0.6 to 3.7 kg. We found that the BAAM consistently produced the desired whistling sound signaling intratracheal placement of the tube in all infants weighing more than 1.5 kg. No adverse effects or complications were noted.
我们评估了贝克气道气流监测仪(BAAM),气流通过该监测仪会发出啸声,以确认气管内导管是否在气管内,评估其安全性和有效性。我们研究了46名体重在0.6至3.7千克之间的新生儿。我们发现,对于所有体重超过1.5千克的婴儿,BAAM始终能发出所需的啸声,表明导管已置于气管内。未观察到不良反应或并发症。