钛涂层聚醚醚酮牙科植入物的初步评估。

Preliminary evaluation of titanium-coated PEEK dental implants.

作者信息

Cook S D, Rust-Dawicki A M

机构信息

Department of Orthopaedic Surgery, Tulane Medical School, New Orleans, Louisiana 70112, USA.

出版信息

J Oral Implantol. 1995;21(3):176-81.

PMID:8699511
Abstract

A total of 40 titanium-coated and uncoated polyetheretherketone (PEEK) cylindrical implants was placed in unicortical sites in the femurs of four mongrel dogs. The implants were evaluated mechanically and histologically at periods of four and eight weeks. Interface attachment strength was determined for each specimen, as were measurements of bone contact, porosity, bone ingrowth, inflammatory response, and mode of failure. Statistical analysis revealed that the uncoated implants had a significantly higher interface shear strength at four weeks, but that there was no significant difference from the coated implants at eight weeks. At both four and eight weeks, the titanium-coated specimens had significantly higher percentages of bone contact. There was no significant increase in bone contact between four and eight weeks for either the coated or uncoated specimens.

摘要

总共40个钛涂层和未涂层的聚醚醚酮(PEEK)圆柱形植入物被放置在4只杂种狗股骨的单皮质部位。在4周和8周时对植入物进行力学和组织学评估。测定每个标本的界面附着强度,以及骨接触、孔隙率、骨长入、炎症反应和失效模式的测量值。统计分析显示,未涂层植入物在4周时具有显著更高的界面剪切强度,但在8周时与涂层植入物没有显著差异。在4周和8周时,钛涂层标本的骨接触百分比均显著更高。涂层或未涂层标本在4周和8周之间骨接触均无显著增加。