A novel, noninvasive method for measuring peripheral venous oxygen saturation (Svo2) in newborn infants using near infrared spectroscopy (NIRS) with venous occlusion is described, along with a comparison with co-oximetry of peripheral venous blood. Peripheral Svo2 was measured in 16 newborn infants by NIRS with venous occlusion, and the results were compared with those obtained by co-oximetry of blood from a forearm vein. There was a significant correlation between the two measurements (r = 0.85, p < 0.0001) and close agreement between the two techniques (mean difference (co-oximeter-NIRS) 6%, limits of agreement-5.1% to 17.1%). Three different methods for analysis of the NIRS data were examined to determine which period during the venous occlusion provided the most accurate measurement. The agreement was the same for all methods of analysis.
描述了一种使用近红外光谱技术(NIRS)并结合静脉闭塞来测量新生儿外周静脉血氧饱和度(Svo2)的新型非侵入性方法,同时与外周静脉血的共血氧测定法进行了比较。通过NIRS结合静脉闭塞对16名新生儿的外周Svo2进行了测量,并将结果与通过前臂静脉血共血氧测定法获得的结果进行了比较。两种测量方法之间存在显著相关性(r = 0.85,p < 0.0001),并且两种技术之间一致性良好(平均差异(共血氧测定仪 - NIRS)为6%,一致性界限为 - 5.1%至17.1%)。研究了三种不同的NIRS数据分析方法,以确定静脉闭塞期间哪个时间段能提供最准确的测量结果。所有分析方法的一致性相同。