Herts B R, Cohen M A, McInroy B, Davros W J, Zepp R C, Einstein D M
Department of Radiology, Cleveland Clinic Foundation, OH 44195, USA.
Radiology. 1996 Sep;200(3):731-5. doi: 10.1148/radiology.200.3.8756923.
To determine the feasibility of use of a power injector to deliver contrast material through central venous catheters for computed tomographic (CT) examinations.
Ioversol 240 and iothalamate meglumine 43% were separately injected through three 9.6-F Hickman catheters and three 10.0-F Leonard catheters with a power injector in an in vitro study. Flow rates of 1.0, 1.5, 2.0, and 2.5 mL/sec were tested. Peak pressures were mechanically recorded from two sites. A 95% prediction interval was calculated for each peak pressure, and the upper limits at the prediction interval were evaluated to determine if it was less than the recommended limit of 25 psi (175 kPa).
Contrast medium, flow rate, and catheter type each statistically significantly affected the measured peak pressures (P = .0001). For each flow rate tested, the upper limits of the prediction interval for the peak pressure at the connection between the coiled tubing and the catheter were below the manufacturer's specified peak pressure.
In vitro analysis demonstrates that power injection of intravenous contrast medium through central venous catheters does not exceed the pressure limits of these catheters at the flow rates tested. In vivo testing to evaluate the safety and efficacy of power injection through central venous catheters is necessary.
确定使用动力注射器通过中心静脉导管注入造影剂进行计算机断层扫描(CT)检查的可行性。
在一项体外研究中,使用动力注射器分别通过三根9.6F的希克曼导管和三根10.0F的伦纳德导管注入碘海醇240和43%的碘他拉酸钠葡甲胺。测试了1.0、1.5、2.0和2.5毫升/秒的流速。从两个部位机械记录峰值压力。计算每个峰值压力的95%预测区间,并评估预测区间的上限,以确定其是否低于推荐的25磅力/平方英寸(175千帕)限值。
造影剂、流速和导管类型均对测量的峰值压力有统计学显著影响(P = .0001)。对于每个测试流速,在盘管与导管连接处的峰值压力预测区间上限均低于制造商规定的峰值压力。
体外分析表明,在所测试的流速下,通过中心静脉导管进行静脉造影剂的动力注射不会超过这些导管的压力限值。有必要进行体内测试以评估通过中心静脉导管进行动力注射的安全性和有效性。