Hollander Scott, Mojibian Hamid, Emery Michael, Tal Michael G
Department of Radiology, Yale-New Haven Hospital, CT 06510, USA.
J Vasc Access. 2012 Jan-Mar;13(1):61-4. doi: 10.5301/JVA.2011.8442.
End-stage renal disease patients with hemodialysis catheters in need of contrast enhanced imaging studies often have limited peripheral venous access. In this study we aimed to determine pressures generated in hemodialysis catheters during power injection of computed tomography (CT) contrast media.
Three different chronic hemodialysis catheters and two acute hemodialysis catheters were included in this study. All catheters were evaluated in vitro. A total volume of 120 cc of CT contrast material was injected at rate of 10 cc/s using a power injector. The catheters were connected to the power injector using a standard connecting tubing. Pressures were simultaneously measured in the power injector as well as in the hemodialysis catheters.
The maximal measured pressures during injection in the power injector averaged 338 PSI (SD ± 8.7 PSI). The maximal measured pressure in the dialysis catheters ranged between 9.17 and 21.2 PSI. Pressures averaged 14.02 PSI (SD ± 3.34 PSI). The average pressure in the power injector was over 23 times higher than the pressure recorded at the hemodialysis catheter. None of the catheters ruptured or deformed during testing.
Pressures measured in hemodialysis catheters during power injection are lower than currently believed and markedly lower than the pressures recorded in the power injector. Standard hemodialysis catheters are likely to be amenable to power contrast injection in hemodialysis patients who have limited venous access. In vivo studies are necessary to confirm these findings.
需要进行增强造影成像检查的终末期肾病血液透析导管患者,其外周静脉通路往往有限。在本研究中,我们旨在确定在计算机断层扫描(CT)造影剂高压注射过程中血液透析导管内产生的压力。
本研究纳入了三种不同的慢性血液透析导管和两种急性血液透析导管。所有导管均在体外进行评估。使用高压注射器以10 cc/s的速率注射总量为120 cc的CT造影剂。导管通过标准连接管与高压注射器相连。在高压注射器以及血液透析导管中同时测量压力。
高压注射器注射过程中测得的最大压力平均为338磅力/平方英寸(标准差±8.7磅力/平方英寸)。透析导管中测得的最大压力在9.17至21.2磅力/平方英寸之间。压力平均为14.02磅力/平方英寸(标准差±3.34磅力/平方英寸)。高压注射器中的平均压力比血液透析导管处记录的压力高出23倍以上。在测试过程中,没有一根导管破裂或变形。
高压注射过程中在血液透析导管中测得的压力低于目前的认知,且明显低于在高压注射器中记录的压力。对于静脉通路有限的血液透析患者,标准血液透析导管可能适合进行高压造影剂注射。需要进行体内研究以证实这些发现。