Nakayama R, Azuma T
Am J Physiol. 1977 Jul;233(1):H168-79. doi: 10.1152/ajpheart.1977.233.1.H168.
A noninvasive method was developed for measuring the digital arterial pressure and the compliance by using a fingertip pneumoplethysmograph and a pneumatic cuff. The compliance (C) of the digital artery was obtained from the peak amplitude of the volume pulse wave (deltaVp-a) under the effect of the cuff pressure (Po-a) by the equation: C = deltavp-a/(Ps - P-o-a) during the dicrotic phase defined in this study. The normal mean value was 11.37 +/- 0.59 X 10(-5) cm3/mmHg. On lowering of the cuff pressure, the moment when the deltaVp value becomes positive is regarded as the systolic pressure (Ps). At the end of the dicrotic phase, the mean amplitude (deltaVss') of the pulse wave during one pulse cycle (SS') and the ratios (deltaVss'/deltaVsd), where deltaVsd is the mean amplitude of the wave during the systolic period (SD), of successive waves after a particular wave fail to increase at the same rate when the cuff pressure decreased below the diastolic pressure. The cuff pressure corresponding to this particular wave is regarded as the diastolic pressure (Pd). The mean value of the mean digital pressure of normotensive subjects was 80.6 +/- 1.2 mmHg.
开发了一种非侵入性方法,通过使用指尖体积描记器和气动袖带测量指动脉压和顺应性。在本研究定义的降中波阶段,根据以下公式从袖带压力(Po-a)作用下的容积脉搏波峰值幅度(deltaVp-a)获得指动脉的顺应性(C):C = deltavp-a/(Ps - P-o-a)。正常平均值为11.37 +/- 0.59 X 10(-5) cm3/mmHg。降低袖带压力时,deltaVp值变为正值的时刻被视为收缩压(Ps)。在降中波阶段结束时,当袖带压力降至舒张压以下时,一个脉搏周期(SS')内脉搏波的平均幅度(deltaVss')以及特定波之后连续波的比率(deltaVss'/deltaVsd)(其中deltaVsd是收缩期(SD)期间波的平均幅度)不会以相同速率增加。与该特定波对应的袖带压力被视为舒张压(Pd)。血压正常受试者的平均指压平均值为80.6 +/- 1.2 mmHg。