Nimura N, Itoh H
Laboratory of Analytical Chemistry, Kitasato University, Tokyo, Japan.
Mol Biotechnol. 1996 Feb;5(1):11-6. doi: 10.1007/BF02762408.
Reversed-phase high-performance liquid chromatographic (RP-HPLC) separation of proteins on chemically bonded silica gel columns is described. Efficiency of nonporous alkylsilyl bonded silica gel is compared with that of a macroporous gel that has been widely used for the purpose. A comparative study of the separation under conventional and fast separation conditions is also given. The fast separation technique on the nonporous reversed-phase column has the advantage of improving the recovery of late-eluting hydrophobic and large proteins, such as ovalbumin and apoferritin.
描述了在化学键合硅胶柱上进行反相高效液相色谱(RP-HPLC)分离蛋白质的方法。将无孔烷基硅烷键合硅胶的分离效率与一种已广泛用于此目的的大孔凝胶的效率进行了比较。还给出了在常规和快速分离条件下分离的对比研究。无孔反相柱上的快速分离技术具有提高后期洗脱的疏水和大蛋白质(如卵清蛋白和脱铁铁蛋白)回收率的优点。