[使用眼前节分析系统原位测量人工晶状体位置的可重复性]

[Reproducibility of measurement of intraocular lens location in situ with an anterior eye segment analysis system].

作者信息

Hirokane K, Kimura W, Kimura T, Sawada T, Ohte A

机构信息

Hikari-kai Kimura Eye and Internal Medicine Hospital, Japan.

出版信息

Nippon Ganka Gakkai Zasshi. 1996 Mar;100(3):240-5.

PMID:8900591
Abstract

The reproducibility of the measurement (accuracy of repeated measurement) of intraocular lens (IOL) location in situ was studied using an anterior eye segment analysis system (NIDEK EAS-1000). The measurements were performed as follows: (1) errors at photography: 1) errors by the individual photographer 2) errors among three photographers. (2) errors at analysis: 1) errors by individual analysists 2) errors among six analysists. The greatest standard deviation was the accuracy of repeated measurement. Accuracy of repeated measurement (tilt and decentration) was (1) 1) 1.104 degrees, 0.083 mm, (1) 2) 0.866 degrees, 0.072 mm, (2) 1) 0.597 degrees, 0.079 mm, (2) 2) 0.562 degrees, 0.079 mm. From the data obtained in this study, the reproducibility of the EAS-1000 was satisfactory for measurement of IOL location.

摘要

使用眼前节分析系统(NIDEK EAS - 1000)研究了人工晶状体(IOL)在位测量的重复性(重复测量的准确性)。测量按以下方式进行:(1)摄影时的误差:1)个体摄影师的误差2)三位摄影师之间的误差。(2)分析时的误差:1)个体分析人员的误差2)六位分析人员之间的误差。最大标准差即为重复测量的准确性。重复测量的准确性(倾斜和偏心)为(1)1)1.104度,0.083毫米,(1)2)0.866度,0.072毫米,(2)1)0.597度,0.079毫米,(2)2)0.562度,0.079毫米。从本研究获得的数据来看,EAS - 1000在测量IOL位置方面的重复性令人满意。