通过门控正电子发射断层扫描术对左心室容积和功能进行无创测量。

Non-invasive measurement of left ventricular volumes and function by gated positron emission tomography.

作者信息

Boyd H L, Gunn R N, Marinho N V, Karwatowski S P, Bailey D L, Costa D C, Camici P G

机构信息

Cyclotron Unit, MRC Clinical Sciences Centre and RPMS, Hammersmith Hospital, London, UK.

出版信息

Eur J Nucl Med. 1996 Dec;23(12):1594-602. doi: 10.1007/BF01249622.

Abstract

To date cardiac positron emission tomography (PET) studies have focussed on the measurement of myocardial blood flow, metabolism and receptors while left ventricular (LV) function and dimensions have been derived from other modalities. The main drawback of this approach is the difficulty of data co-registration, which limits clinical interpretation. The aim of this study was to evaluate whether it is possible to measure absolute cardiac volumes, and consequently LV function parameters such as ejection fraction, and wall motion with gated PET. Nineteen patients underwent a PET scan and planar radionuclide ventriculography (MUGA) within 9+/-9 days. A 9-min scan (16 gates/cardiac cycle) was acquired after inhalation of 3 MBq/ml of oxygen-15 labelled carbon monoxide at the rate of 500 ml/min over 4 min using a multislice PET camera. Noise reduction was performed on the gated image to enhance the definition of the ventricles before reslicing to the short-axis view. A threshold value was used to detect the edge of the LV at each gate. LV volumes at each gate were estimated by summing the volume of voxels within the LV boundary. PET measurements of LV volumes were as follows: LV end-diastolic volume ranged from 72 to 233 ml and LV end-systolic volume ranged from 24 to 203 ml. Phantom experiments supported the validity of this approach for estimating volumes. LV ejection fraction measured with MUGA was 38.4%+/-16.3% (range 15%-71%) and that measured with PET was 39.6%+/-17.7% (range 9%-72%) (P=NS). The LV ejection fraction measurements were highly correlated (r2=0.824). These results indicate that: (1) absolute end-diastolic and end-systolic volumes can be quantified using gated PET and (2) LV ejection fraction can be accurately measured by gated PET simultaneously with the other physiological PET parameters.

摘要

迄今为止,心脏正电子发射断层扫描(PET)研究主要集中在心肌血流、代谢和受体的测量上,而左心室(LV)功能和尺寸则通过其他方式得出。这种方法的主要缺点是数据配准困难,这限制了临床解读。本研究的目的是评估是否可以通过门控PET测量绝对心脏容积,进而测量诸如射血分数等左心室功能参数以及室壁运动。19名患者在9±9天内接受了PET扫描和平面放射性核素心室造影(MUGA)。使用多层PET相机,在以500 ml/min的速率吸入3 MBq/ml的氧-15标记一氧化碳4分钟后,进行一次9分钟的扫描(16个门/心动周期)。在门控图像上进行降噪处理,以增强心室的清晰度,然后再重切成短轴视图。使用阈值检测每个门控时左心室的边缘。通过将左心室内边界内体素的体积相加来估计每个门控时的左心室容积。左心室容积的PET测量结果如下:左心室舒张末期容积范围为72至233 ml,左心室收缩末期容积范围为24至203 ml。模型实验支持了这种容积估计方法的有效性。用MUGA测量的左心室射血分数为38.4%±16.3%(范围15%-71%),用PET测量的为39.6%±17.7%(范围9%-72%)(P=无显著性差异)。左心室射血分数测量结果高度相关(r2=0.824)。这些结果表明:(1)可以使用门控PET对舒张末期和收缩末期绝对容积进行定量,(2)门控PET可以在测量其他生理PET参数的同时准确测量左心室射血分数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验