Elliott M R, Thrush A J
Department of Medical Physics, Leicester Royal Infirmary, UK.
Physiol Meas. 1996 Nov;17(4):259-65. doi: 10.1088/0967-3334/17/4/003.
Detailed images of arteries can be formed by inserting intravascular ultrasound (IVUS) catheters within them. These catheters contain rotating ultrasound transducers and are capable of displaying disease and features such as the layers of artery walls. There is, however, relatively little information available on IVUS resolution even though this is necessary to interpret the images. Techniques for assessing IVUS resolution in three dimensions are described for a 20 MHz Boston Scientific catheter and a Diasonics IVUS scanner as well as the results obtained. Out-of-plane resolution was measured by finding the ultrasound beam thickness. Axial and lateral resolution were assessed by the minimum separation of two point targets when their images could just be distinguished. The out-of-plane resolution was found to improve slightly with distance from the catheter, varying from 2.0-1.5 mm. Lateral resolution worsened in a linear manner with distance. At a distance of 5 mm from the catheter, lateral resolution was 0.83 mm. The axial resolution was the best of the three resolutions and was actually limited by the screen display scale. With the highest magnification the axial resolution was approximately 0.19 mm and was constant with distance.
通过将血管内超声(IVUS)导管插入动脉内,可以形成动脉的详细图像。这些导管包含旋转超声换能器,能够显示疾病以及诸如动脉壁各层等特征。然而,尽管这对于解释图像是必要的,但关于IVUS分辨率的可用信息相对较少。本文描述了用于评估20 MHz波士顿科学公司导管和迪阿松尼斯IVUS扫描仪三维IVUS分辨率的技术以及获得的结果。通过确定超声束厚度来测量平面外分辨率。当两点目标的图像刚好能够区分时,通过两点目标的最小间距来评估轴向分辨率和横向分辨率。发现平面外分辨率随着与导管距离的增加而略有提高,范围从2.0 - 1.5毫米不等。横向分辨率随着距离呈线性恶化。在距离导管5毫米处,横向分辨率为0.83毫米。轴向分辨率是三种分辨率中最好的,实际上受屏幕显示比例限制。在最高放大倍数下,轴向分辨率约为0.19毫米,并且随距离保持恒定。