Suppr超能文献

[Pruritometer 1: Portable measuring system for quantifying scratching as an objective measure of cholestatic pruritus].

作者信息

Stein H, Bijak M, Heerd E, Mayr W, Müller C, Rafolt D, Schnetz G, Unger E

机构信息

Institut für Biomedizinische Technik und Physik, Universität Wien.

出版信息

Biomed Tech (Berl). 1996 Sep;41(9):248-52. doi: 10.1515/bmte.1996.41.9.248.

Abstract

Objective assessment of subjective symptoms such as pruritus always presents problems, which can often be resolved only indirectly. The objectification of pruritus was necessitated by a study on the efficacy of a serotonin antagonist used as treatment of cholestatic pruritus. In the present paper, a portable measuring system for the indirect objective assessment of pruritus via the quantification of scratching is described. A piezoelectric scratch-vibration sensor for attachment to the middle finger of the patient's dominant hand was developed. A sensor interface detects the scratching signals and generates pulses that are then summed in an adapted sports watch. The entire system-Pruritometer 1- is worn by the patient like a wrist watch, and is characterized by ease of handling. Acceptance by the patient is reported to be good. A statistically good correlation between measured (Pruritometer) and visually counted scratches was demonstrated. Pruritometer 2, which will enable scratch frequency and intensity distribution over time to be determined, is presently being developed.