[Quality control of mammography: potential use of a new device for testing the image quality and the dose].

作者信息

Di Maggio C, Gennaro G, Gambaccini M, Cattozzo S

机构信息

Istituto di Radiologia, Università di Padova.

出版信息

Radiol Med. 1996 Sep;92(3):261-6.

PMID:8975313
Abstract

The authors tested a new equipment containing a phantom to check quality mammographic image and an electronic instrument to measure incident radiation (Phan-EX). The phantom contains objects which have similar characteristics to the details of diagnostic interest and allows an easy and accurate estimation of image quality on the basis of the number of objects detected. The radiation detector, consisting of a solid state photodiode, provides an integer proportional to the input exposure. The results obtained highlighted a good reproducibility of instrument output and a good relationship between these values and the exposure ones obtained with a much more expensive ionization chamber. The device response was linear in varying both the anodic exposure (mAs) and the x-ray tube voltage (varying the energy, the calibration factor varied only by a small percent). The authors conclude that the Phan-EX may be useful for quality assurance of the mammographic unit. Furthermore, they suggest some of its practical applications: daily tests of qualitative of the mammographic unit and the treatment system performance, control of the automatic exposure probe, choice and optimization of exposure and treatment of the film parameters, the determination of the output of the x-ray tube, the evaluation of the sensitivity of different screen/film systems.

摘要

相似文献

1
[Quality control of mammography: potential use of a new device for testing the image quality and the dose].
Radiol Med. 1996 Sep;92(3):261-6.
2
Radiographic quality control devices.射线照相质量控制设备。
Health Devices. 2000 Apr;29(4):97-139.
3
[The procedures and results of a quality control program in mammography carried out on a regional basis].
Radiol Med. 1996 Mar;91(3):187-93.
4
Mammographic equipment, technique, and quality control.
Curr Opin Radiol. 1991 Aug;3(4):571-8.
5
Quality assurance in screening mammography.乳腺钼靶筛查中的质量保证
Health Devices. 1990 May-Jun;19(5-6):152-98.
6
Mammography equipment: principles, features, selection.
Radiol Clin North Am. 1987 Sep;25(5):897-911.
7
Monochromatic x-rays in digital mammography.数字乳腺摄影中的单色X射线。
Invest Radiol. 2005 Jan;40(1):33-9.
8
The effect of a compression paddle on energy response, calibration and measurement with mammographic dosimeters using ionization chambers and solid-state detectors.压迫板对使用电离室和固态探测器的乳腺摄影剂量仪的能量响应、校准及测量的影响。
Phys Med Biol. 2009 Feb 21;54(4):1047-59. doi: 10.1088/0031-9155/54/4/015. Epub 2009 Jan 23.
9
Reference levels for image quality in mammography.
Radiat Prot Dosimetry. 2008;129(1-3):170-2. doi: 10.1093/rpd/ncn080. Epub 2008 Mar 28.
10
[Experimental investigations for dose reduction by optimizing the radiation quality for digital mammography with an a-Se detector].[利用非晶硒探测器优化数字乳腺摄影的辐射质量以降低剂量的实验研究]
Rofo. 2007 May;179(5):487-91. doi: 10.1055/s-2007-962983.