Comparison of the Omron HEM-713C automated blood pressure monitor with a standard ausculatory method using a mercury manometer.

作者信息

Mufunda J, Sparks B, Chifamba J, Dakwa C, Matenga J A, Adams J M, Sparks H V

机构信息

Department of Physiology, University of Zimbabwe, Harare, Zimbabwe.

出版信息

Cent Afr J Med. 1996 Aug;42(8):230-2.

PMID:8990566
Abstract

OBJECTIVES

To compare the Omron HEM-713C automated blood pressure machine with the standard ausculatory method using a mercury manometer.

DESIGN

Blood pressures of randomly selected subjects were measured using both the Omron HEM-713C and the mercury manometer.

SETTING

Dombotombo surburb in Marondera, Zimbabwe.

SUBJECTS

One hundred and sixteen subjects 25 years and above (47 males and 69 females) randomly selected in Marondera.

MAIN OUTCOME MEASURE

Systolic blood pressure and diastolic blood pressure.

RESULTS

The Omron HEM-713C passed with a grade B for both systolic and diastolic blood pressures when using the British Hypertension Society protocol. It also passed both systolic and diastolic criteria for Association of the Advancement of Medical Instrumentation.

CONCLUSION

The Omron HEM-713C compares well with the standard mercury manometer, we therefore recommend its use in both research and clinical applications which require blood pressure measurements.