Swee R G, Gray J E, Beabout J W, McLeod R A, Cooper K L, Bond J R, Wenger D E
Department of Diagnostic Radiology, Mayo Clinic and Foundation, Rochester, MN 55905, USA.
AJR Am J Roentgenol. 1997 Feb;168(2):539-42. doi: 10.2214/ajr.168.2.9016243.
Computed radiography of the musculoskeletal system has the potential to become a powerful tool in the practice of diagnostic radiology. It addresses many of the geographic and film-distribution concerns facing diagnostic imaging. We undertook this study to compare and document the quality of computed radiographs and conventional screen-film images before widespread implementation.
We evaluated clinical images using direct comparison. Bilateral hand images from 50 patients were scored independently by six musculoskeletal radiologists. In each case one hand was imaged with a conventional screen-film technique and the other with computed radiography. Images were masked to eliminate as much bias as possible. The numeric scores assigned to the images by the observers were analyzed using Student's t test.
Computed radiographs were judged with statistical significance to be better than conventional screen-film images in all features judged by the observers, including bone cortex, bone trabeculae, corticomedullary junction, distal phalangeal tuft, soft tissues, fat planes, bone-soft-tissue interface, and overall contrast and density.
The statistically significant determination that the image quality of computed radiographs is at least as good as screen-film images allows confident use of computed radiography and enables radiologists to take advantage of its many other practical capabilities related to image distribution, storage, cost, and geographic coverage without sacrificing image quality.
计算机X线摄影在肌肉骨骼系统中有可能成为诊断放射学实践中的一项强大工具。它解决了诊断成像面临的许多地理和胶片分发方面的问题。我们开展这项研究是为了在广泛应用之前比较并记录计算机X线片和传统屏-片图像的质量。
我们采用直接比较的方法评估临床图像。50例患者的双侧手部图像由6名肌肉骨骼放射科医生独立评分。在每种情况下,一只手采用传统屏-片技术成像,另一只手采用计算机X线摄影成像。对图像进行遮盖以尽可能消除偏差。使用学生t检验分析观察者给图像赋予的数值评分。
在观察者判断的所有特征方面,包括骨皮质、骨小梁、皮质髓质交界、远节指骨末端、软组织、脂肪平面、骨-软组织界面以及整体对比度和密度,计算机X线片在统计学上被判定优于传统屏-片图像。
计算机X线片图像质量在统计学上显著优于屏-片图像这一判定结果,使得可以放心使用计算机X线摄影,并且让放射科医生能够利用其在图像分发、存储、成本和地理覆盖范围等方面的许多其他实用功能,而不牺牲图像质量。