Jones A, Ansell C, Jerrom C, Honey I D
1 Medical Physics Department, Guy's and St Thomas' NHS Foundation Trust, London, UK.
Br J Radiol. 2015 Jun;88(1050):20140660. doi: 10.1259/bjr.20140660. Epub 2015 Mar 27.
The purpose of this study was to evaluate the effect of beam quality on the image quality (IQ) of ankle radiographs of paediatric patients in the age range of 0-1 year whilst maintaining constant effective dose (ED).
Lateral ankle radiographs of an infant foot phantom were taken at a range of tube potentials (40.0-64.5 kVp) with and without 0.1-mm copper (Cu) filtration using a Trixell Pixium 4600 detector (Trixell, Morains, France). ED to the patient was computed for the default exposure parameters using PCXMC v. 2.0 and was fixed for other beam qualities by modulating the tube current-time product. The contrast-to-noise ratio (CNR) was measured between the tibia and adjacent soft tissue. The IQ of the phantom images was assessed by three radiologists and a reporting radiographer. Four IQ criteria were defined each with a scale of 1-3, giving a maximum score of 12. Finally, a service audit of clinical images at the default and optimum beam qualities was undertaken.
The measured CNR for the 40 kVp/no Cu image was 12.0 compared with 7.6 for the default mode (55 0.1 mm Cu). An improvement in the clinical IQ scores was also apparent at this lower beam quality.
Lowering tube potential and removing filtration improved the clinical IQ of paediatric ankle radiographs in this age range.
There are currently no UK guidelines on exposure protocols for paediatric imaging using direct digital radiography. A lower beam quality will produce better IQ with no additional dose penalty for infant extremity imaging.
本研究旨在评估在保持有效剂量(ED)恒定的情况下,射线质对0至1岁儿科患者踝关节X线片图像质量(IQ)的影响。
使用Trixell Pixium 4600探测器(法国莫兰的Trixell公司),在一系列管电压(40.0 - 64.5 kVp)下,对婴儿足部模型进行踝关节侧位X线摄影,分别采用有和没有0.1毫米铜(Cu)滤过的情况。使用PCXMC v. 2.0根据默认曝光参数计算患者的ED,并通过调节管电流 - 时间乘积对其他射线质固定该值。测量胫骨与相邻软组织之间的对比噪声比(CNR)。由三名放射科医生和一名报告放射技师评估模型图像的IQ。定义了四个IQ标准,每个标准的评分范围为1至3,最高得分为12分。最后,对默认和最佳射线质下的临床图像进行了服务审核。
40 kVp/无Cu图像的测量CNR为12.0,而默认模式(55 kVp/0.1毫米Cu)为7.6。在这种较低的射线质下,临床IQ评分也有明显改善。
降低管电压并去除滤过可改善该年龄范围内儿科踝关节X线片的临床IQ。
目前英国尚无关于使用直接数字化X线摄影进行儿科成像的曝光方案指南。较低的射线质将产生更好的IQ,且对婴儿四肢成像不会增加额外剂量。