跟骨的高分辨率磁共振成像:小梁结构的年龄相关性变化以及与双能X线吸收测定法测量结果的比较

High resolution magnetic resonance imaging of the calcaneus: age-related changes in trabecular structure and comparison with dual X-ray absorptiometry measurements.

作者信息

Ouyang X, Selby K, Lang P, Engelke K, Klifa C, Fan B, Zucconi F, Hottya G, Chen M, Majumdar S, Genant H K

机构信息

Department of Radiology, University of California, San Francisco 94143, USA.

出版信息

Calcif Tissue Int. 1997 Feb;60(2):139-47. doi: 10.1007/s002239900204.

Abstract

A high-resolution magnetic resonance imaging (MRI) protocol, together with specialized image processing techniques, was applied to the quantitative measurement of age-related changes in calcaneal trabecular structure. The reproducibility of the technique was assessed and the annual rates of change for several trabecular structure parameters were measured. The MR-derived trabecular parameters were compared with calcaneal bone mineral density (BMD), measured by dual X-ray absorptiometry (DXA) in the same subjects. Sagittal MR images were acquired at 1.5 T in 23 healthy women (mean age: 49.3 +/- 16.6 [SD]), using a three-dimensional gradient echo sequence. Image analysis procedures included internal gray-scale calibration, bone and marrow segmentation, and run-length methods. Three trabecular structure parameters, apparent bone volume (ABV/TV), intercept thickness (I.Th), and intercept separation (I.Sp) were calculated from the MR images. The short- and long-term precision errors (mean %CV) of these measured parameters were in the ranges 1-2% and 3-6%, respectively. Linear regression of the trabecular structure parameters vs. age showed significant correlation: ABV/TV (r2 = 33.7%, P < 0.0037), I.Th (r2 = 26.6%, P < 0.0118), I.Sp (r2 = 28.9%, P < 0.0081). These trends with age were also expressed as annual rates of change: ABV/TV (-0.52%/year), I.Th (-0.33%/year), and I.Sp (0.59%/year). Linear regression analysis also showed significant correlation between the MR-derived trabecular structure parameters and calcaneal BMD values. Although a larger group of subjects is needed to better define the age-related changes in trabecular structure parameters and their relation to BMD, these preliminary results demonstrate that high-resolution MRI may potentially be useful for the quantitative assessment of trabecular structure.

摘要

一种高分辨率磁共振成像(MRI)方案,连同专门的图像处理技术,被应用于跟骨小梁结构中与年龄相关变化的定量测量。评估了该技术的可重复性,并测量了几个小梁结构参数的年变化率。将磁共振衍生的小梁参数与通过双能X线吸收法(DXA)在同一受试者中测量的跟骨骨密度(BMD)进行比较。在1.5 T下,使用三维梯度回波序列,对23名健康女性(平均年龄:49.3±16.6 [标准差])采集矢状面MR图像。图像分析程序包括内部灰度校准、骨和骨髓分割以及游程长度法。从MR图像中计算出三个小梁结构参数,表观骨体积(ABV/TV)、截距厚度(I.Th)和截距间距(I.Sp)。这些测量参数的短期和长期精度误差(平均%CV)分别在1 - 2%和3 - 6%的范围内。小梁结构参数与年龄的线性回归显示出显著相关性:ABV/TV(r2 = 33.7%,P < 0.0037),I.Th(r2 = 26.6%,P < 0.0118),I.Sp(r2 = 28.9%,P < 0.0081)。这些随年龄的变化趋势也以年变化率表示:ABV/TV(-0.52%/年),I.Th(-0.33%/年),I.Sp(0.59%/年)。线性回归分析还显示磁共振衍生的小梁结构参数与跟骨BMD值之间存在显著相关性。尽管需要更大规模的受试者群体来更好地定义小梁结构参数与年龄相关的变化及其与BMD的关系,但这些初步结果表明,高分辨率MRI可能对小梁结构的定量评估有潜在用途。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验